Literature DB >> 17545430

Total elbow prosthesis loosening caused by ulnar component pistoning.

Emilie V Cheung1, Shawn W O'Driscoll.   

Abstract

BACKGROUND: Linked semiconstrained total elbow prostheses have been used successfully but may be at higher risk for implant loosening than unlinked implants are. The purpose of the present report was to describe a previously unreported and potentially preventable cause of mechanical loosening of the ulnar component of a linked total elbow prosthesis.
METHODS: A series of ten patients who had painful pistoning of the polymethylmethacrylate-coated ulnar component of a Coonrad-Morrey linked total elbow prosthesis were evaluated clinically and radiographically.
RESULTS: All ten patients complained of elbow pain, and eight had a distinct sensation of the ulnar component moving within the ulna. Six patients either complained of squeaking within the elbow or could demonstrate squeaking on examination. Four patients had a complete radiolucent line around the ulnar component or the cement mantle, and six had an incomplete line around the ulnar component. Six patients had a radiolucent gap between the cement and the tip of the ulnar prosthesis. Two patients had proximal migration of the ulnar component within the cement mantle on lateral flexion radiographs. Three patients had anterior impingement, such as between the anterior flange of the humeral implant and a prominent coronoid process, on lateral flexion radiographs. At the time of revision arthroplasty, all ten patients were found to have a loose ulnar component, which was successfully revised with or without impaction grafting. At the time of the most recent follow-up, nine of the ten ulnar components were intact and stable. Three patients required an additional reoperation: one required triceps repair, one required revision of a loose humeral component, and one required a revision total elbow arthroplasty.
CONCLUSIONS: Pistoning of the ulnar component in the cement mantle leading to failure by means of a pullout mechanism can occur in association with the Coonrad-Morrey total elbow prosthesis with a polymethylmethacrylate-precoated ulnar component. To prevent this problem following any total elbow arthroplasty, the surgeon should check for anterior impingement intraoperatively by ensuring that there is no contact between the anterior flange and a prominent coronoid process or the cement and that no distraction of the trial ulnar component from the ulna occurs with passive elbow flexion. This condition also can be avoided by ensuring that the ulnar component is not inserted too far distally. This mechanism of failure should be considered when future total elbow arthroplasty implants are designed. LEVEL OF EVIDENCE: Therapeutic Level IV.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17545430     DOI: 10.2106/JBJS.F.00376

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  13 in total

Review 1.  Triceps on approach for total elbow arthroplasty: worth preserving? A review of approaches for total elbow arthroplasty.

Authors:  Simon J Booker; Chris D Smith
Journal:  Shoulder Elbow       Date:  2016-12-13

2.  Linked semi-constrained total elbow prosthesis in chronic arthritis: results of 18 cases.

Authors:  Maurizio Corradi; Marco Frattini; Bruno Panno; Silvio Tocco; Francesco Pogliacomi
Journal:  Musculoskelet Surg       Date:  2010-05

Review 3.  [Allograft reconstruction in revision elbow arthroplasty].

Authors:  F Moro
Journal:  Orthopade       Date:  2017-12       Impact factor: 1.087

4.  Clinical results of revision total elbow arthroplasty: comparison of infected and non-infected total elbow arthroplasty.

Authors:  Jae-Man Kwak; Erica Kholinne; Yucheng Sun; Myung-Sun Kim; Kyoung-Hwan Koh; In-Ho Jeon
Journal:  Int Orthop       Date:  2019-01-07       Impact factor: 3.075

Review 5.  Distal humeral hemiarthroplasty: indications, results, and complications. A systematic review.

Authors:  John Dunn; Nicholas Kusnezov; Miguel Pirela-Cruz
Journal:  Hand (N Y)       Date:  2014-12

6.  Primary elbow arthroplasty: problems and solutions.

Authors:  Joaquin Sanchez-Sotelo
Journal:  Shoulder Elbow       Date:  2016-11-16

7.  Unexpected high early failure rate of the Nexel total elbow arthroplasty.

Authors:  Mark E Morrey; Chad Songy; Jacob J Triplet; Adnan N Cheema; Shawn W O'Driscoll; Joaquin Sanchez-Sotelo; Bernard F Morrey
Journal:  JSES Int       Date:  2022-05-06

8.  Clinical outcomes and complications following primary total elbow arthroplasty using the Latitude prosthesis.

Authors:  David Cinats; Aaron J Bois; Kevin A Hildebrand
Journal:  Shoulder Elbow       Date:  2018-05-01

9.  Does the triceps-on approach affect alignment in total elbow arthroplasty? A cadaveric study.

Authors:  Andrew King; Jonathan P Evans; Simon J Booker; James Cs Beazley; Robin Js Jones; William Jc Thomas; Christopher Smith
Journal:  Shoulder Elbow       Date:  2018-05-25

Review 10.  Revision total elbow replacement.

Authors:  Kuen Chin; Simon Lambert
Journal:  J Clin Orthop Trauma       Date:  2021-07-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.