Literature DB >> 17538648

Optimization of 6-carboxy-X-rhodamine concentration for real-time polymerase chain reaction using molecular beacon chemistry.

Gehua Wang1, Erin Becker, Christine Mesa.   

Abstract

The optimal 6-carboxy-X-rhodamine (ROX) concentration, which is used as a passive reference dye for real-time quantitative polymerase chain reaction (PCR) with molecular beacon chemistry, was determined with the Mx4000 Multiplex Quantitative PCR System. Additionally, the effects of changing ROX concentrations on PCR reproducibility, Ct values, and efficiency were investigated with this system by using the PCR data obtained from amplification of the Escherichia coli shiga toxin 2 (stx2) gene and the Campylobacter jejuni luxS gene. This study indicated that different ROX concentrations influence many aspects of the real-time PCR reaction. ROX concentration variation could have consequences in the analysis of quantitative data and may lead to erroneous results. This study further indicated that the optimal ROX concentration is 60 nmol/L for real-time PCR, using molecular beacon chemistry for PCR assay of luxS and stx2 genes.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17538648     DOI: 10.1139/W06-145

Source DB:  PubMed          Journal:  Can J Microbiol        ISSN: 0008-4166            Impact factor:   2.419