Rectum dose analysis employing a multi-purpose brachytherapy phantom.

HyunDo Huh1, WooChul Kim, John J K Loh, Suk Lee, Chul Yong Kim, SangHoon Lee, DongOh Shin, Dongho Shin, SamJu Cho, Jisun Jang, Sangwook Lim, Kwang Hwan Cho, Sooil Kwon, SeongHoon Kim.   

Abstract

PURPOSE: It is difficult to reproduce a brachytherapy measurement because of changes in the rectal shape during inter-fraction. We constructed a multi-purpose brachytherapy phantom (MPBP) and reproduced the same conditions found in actual therapy. We further attempted to apply the measured optimal dose to reduce rectal complications.
METHODS: A measured dose was administered at rectal reference point R1 using a diode detector in four patients who used a tandem and ovoid in brachytherapy for carcinoma of the cervix. A total number of 20 rectal dose measurements were performed five times per patient. In addition, discrepancies in the set-up of the diode detector were analyzed with each repetitive measurement. After reproducing the same conditions as found in actual therapy using a multi-function applicator (MFA) in the multi-purpose brachytherapy phantom constructed for this study, the dose was measured at reference points in the rectum using a thermoluminescence dosimeter (TLD).
RESULTS: According to the discrepancies measured in the set-up using a diode detector, Patient 1 showed a maximum value of 11.25 +/- 0.95 mm in the Y direction, Patients 2 and 3 exhibited 9.90 +/- 2.40 mm and 20.85 +/- 4.50 mm in the Z direction, respectively. Patient 4 showed 19.15 +/- 3.33 mm in the Z direction. In addition, values of the mean dose according to the position of the diode detector were recorded as 122.82 +/- 7.96-323.78 +/- 11.16 cGy. In the measured results for TLD in an MPBP, relative error for Patients 1 and 4 at the rectal reference point R2 were a maximum of 8.6 and 7.7%, respectively. For Patients 2 and 3 they were 1.7 and 1.2%, respectively. Furthermore, the dose measured at point R1 and R2 exhibited values approximately 1.7-8.6% higher than the dose calculated in advance, excluding point R1 in Patient 2. The discrepancies in the set-up owing to repetitive measurements and alterations in dosage according to these changes were not analyzed. It was evident that the relative error between the calculated and measured value was within 15%, which was allowable according to the recommendations by the American Association of Physicists in Medicine (AAPM).
CONCLUSIONS: The multi-purpose brachytherapy phantom constructed for this study successfully reproduced an optimal dose measured under the same conditions found in actual therapy in which the dose was precisely analyzed at a rectal reference point. In addition, these results were considered reliable and applicable for dose optimization before applying therapy using the measured data from the phantom in order to reduce rectal complications.

Entities:  

Mesh:

Year:  2007        PMID: 17538010     DOI: 10.1093/jjco/hym032

Source DB:  PubMed          Journal:  Jpn J Clin Oncol        ISSN: 0368-2811            Impact factor:   3.019


  4 in total

1.  In vivo dosimetry with semiconductors in medium dose rate (MDR) brachytherapy for cervical cancer.

Authors:  Mahmoud Allahverdi; Ramin Jaberi; Mehdi Aghili; Fatemeh Ghahremani; Ghazale Geraily
Journal:  Jpn J Radiol       Date:  2012-12-05       Impact factor: 2.374

2.  HDR Brachytherapy Dose Distribution is Influenced by the Metal Material of the Applicator.

Authors:  Chin-Hui Wu; Yi-Jen Liao; An-Cheng Shiau; Hsin-Yu Lin; Yen-Wan Hsueh Liu; Shih-Ming Hsu
Journal:  Sci Rep       Date:  2015-12-11       Impact factor: 4.379

Review 3.  Comparative Analysis of 60Co and 192Ir Sources in High Dose Rate Brachytherapy for Cervical Cancer.

Authors:  Aiping Wen; Xianliang Wang; Bingjie Wang; Chuanjun Yan; Jingyue Luo; Pei Wang; Jie Li
Journal:  Cancers (Basel)       Date:  2022-09-29       Impact factor: 6.575

4.  The effect of tandem-ovoid titanium applicator on points A, B, bladder, and rectum doses in gynecological brachytherapy using 192Ir.

Authors:  Mohammad Hosein Sadeghi; Sedigheh Sina; Amir Mehdizadeh; Reza Faghihi; Vahed Moharramzadeh; Ali Soleimani Meigooni
Journal:  J Contemp Brachytherapy       Date:  2018-02-28