Literature DB >> 17520947

A mechanical evaluation of implants placed with different surgical techniques into the trabecular bone of goats.

Manal M Shalabi1, Johannes G C Wolke, Anja J E de Ruijter, John A Jansen.   

Abstract

The aim of the study was to assess the effects of surgical technique and implant surface roughness on implant fixation. A total of 48 screw implants with machined or etched surface topographies were placed into the femoral condyles of goats. The implant sites were prepared by a conventional technique, by undersized preparation, or by the osteotome technique. Bone tissue responses were evaluated after 12 weeks of healing by removal torque testing and histologic analysis using scanning electron microscope. The cumulative removal torque value of the etched implants placed with the undersized technique (98 +/- 29 Ncm) was higher (50 +/- 35 Ncm) to a level of statistical significance than machined surface implants placed by the osteotome technique. Scanning electron microscope evaluation indicated that all implants showed interfacial bone contact. The torque test resulted in fracture at the bone-implant interface for all experimental conditions. Installation of etched implants using an undersized preparation of the implant bed resulted in superior bonding strength with the surrounding bone at 12 weeks after surgery. Evidently, the undersized preparation technique improved the early fixation of oral implants in this study.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17520947     DOI: 10.1563/0-827.1

Source DB:  PubMed          Journal:  J Oral Implantol        ISSN: 0160-6972            Impact factor:   1.779


  3 in total

1.  New Osseodensification Implant Site Preparation Method to Increase Bone Density in Low-Density Bone: In Vivo Evaluation in Sheep.

Authors:  Paolo Trisi; Marco Berardini; Antonello Falco; Michele Podaliri Vulpiani
Journal:  Implant Dent       Date:  2016-02       Impact factor: 2.454

2.  Assessment of Temperature Rise and Time of Alveolar Ridge Splitting by Means of Er:YAG Laser, Piezosurgery, and Surgical Saw: An Ex Vivo Study.

Authors:  Jacek Matys; Rafał Flieger; Marzena Dominiak
Journal:  Biomed Res Int       Date:  2016-11-10       Impact factor: 3.411

Review 3.  Methods to Improve Osseointegration of Dental Implants in Low Quality (Type-IV) Bone: An Overview.

Authors:  Hamdan S Alghamdi
Journal:  J Funct Biomater       Date:  2018-01-13
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.