| Literature DB >> 17520729 |
Albert P Chen1, Charles H Cunningham, Esin Ozturk-Isik, Duan Xu, Ralph E Hurd, Douglas A C Kelley, John M Pauly, John Kurhanewicz, Sarah J Nelson, Daniel B Vigneron.
Abstract
Prostate MR spectroscopic imaging (MRSI) at 3T may provide two-fold higher spatial resolution over 1.5T, but this can result in longer acquisition times to cover the entire gland using conventional phase-encoding. In this study, flyback echo-planar readout trajectories were incorporated into a Malcolm Levitt's composite-pulse decoupling sequence (MLEV)-point-resolved spectroscopy sequence (PRESS) to accelerate the acquisition of large array (16 x 16 x 8), high spatial (0.154 cm(3)) resolution MRSI data by eight-fold to just 8.5 minutes. Artifact free, high-quality MRSI data was obtained in nine prostate cancer patients. Easy data reconstruction and the robustness of the flyback echo-planar encoding make this technique particularly suitable for the clinical setting. The short acquisition time provided by this method reduces the 3T prostate MRI/MRSI exam time, allows longer repetition times, and/or allows the acquisition of additional MR acquisitions within the same exam. (c) 2007 Wiley-Liss, Inc.Entities:
Mesh:
Year: 2007 PMID: 17520729 DOI: 10.1002/jmri.20916
Source DB: PubMed Journal: J Magn Reson Imaging ISSN: 1053-1807 Impact factor: 4.813