Literature DB >> 17518444

Systematic comparison of bias and precision data obtained with multiple-point and one-point calibration in six validated multianalyte assays for quantification of drugs in human plasma.

Frank T Peters1, Hans H Maurer.   

Abstract

One-point (linear through zero) calibration is often used as a compromise between necessary calibration, workload, and time. The aim of the present study was to systematically check the applicability of one-point calibration by comparing bias and precision data obtained with full and one-point calibration. Data from validation studies of six mass spectrometry-based multianalyte bioanalytical assays were used for this purpose. Bias and intermediate precision datasets of full multiple-point calibration were compared to six one-point calibration datasets (A-F in rising calibrator concentration order) calculated from the same raw data. The datasets were statistically compared using the Friedman test followed by Dunn's multiple comparison test. The results obtained with full calibration and the different one-point calibrations were found to differ significantly (P < 0.05) in all of the six studied methods. The best one-point calibration results were obtained with calibrator D, with which acceptance criteria for bias and precision were fulfilled for the majority of analytes. However, some extremely high bias and precision data were obtained for some analytes in the low-concentration range. In conclusion, one-point calibration with a calibrator close to the center of the full calibration range can be a feasible alternative to full calibration.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17518444     DOI: 10.1021/ac070054s

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  3 in total

1.  Simultaneous quantitation of folates, flavins and B6 metabolites in human plasma by LC-MS/MS assay: Applications in colorectal cancer.

Authors:  Isaac Asante; Hua Pei; Eugene Zhou; Siyu Liu; Darryl Chui; EunJeong Yoo; Stan G Louie
Journal:  J Pharm Biomed Anal       Date:  2018-05-22       Impact factor: 3.935

2.  Novel assays for quality evaluation of XueBiJing: Quality variability of a Chinese herbal injection for sepsis management.

Authors:  Xuan Yu; Wei Niu; Ya-Ya Wang; Olajide E Olaleye; Jia-Nan Wang; Meng-Yuan Duan; Jun-Ling Yang; Rong-Rong He; Zi-Xuan Chu; Kai Dong; Gui-Ping Zhang; Chang-Xiao Liu; Chen Cheng; Chuan Li
Journal:  J Pharm Anal       Date:  2022-01-26

3.  Presence of acyl-homoserine lactones in 57 members of the Vibrionaceae family.

Authors:  A A Purohit; J A Johansen; H Hansen; H-K S Leiros; A Kashulin; C Karlsen; A Smalås; P Haugen; N P Willassen
Journal:  J Appl Microbiol       Date:  2013-06-28       Impact factor: 3.772

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.