Literature DB >> 17508238

Initial experience with the resonance metallic stent for antegrade ureteric stenting.

Tze M Wah1, Henry C Irving, Jon Cartledge.   

Abstract

BACKGROUND AND
PURPOSE: We describe our initial experience with a new metallic ureteric stent which has been designed to provide long-term urinary drainage in patients with malignant ureteric strictures. The aim is to achieve longer primary patency rates than conventional polyurethane ureteric stents, where encrustation and compression by malignant masses limit primary patency. The Resonance metallic double-pigtail ureteric stent (Cook, Ireland) is constructed from coiled wire spirals of a corrosion-resistant alloy designed to minimize tissue in-growth and resist encrustation, and the manufacturer recommends interval stent change at 12 months.
METHODS: Seventeen Resonance stents were inserted via an antegrade approach into 15 patients between December 2004 and March 2006. The causes of ureteric obstruction were malignancies of the bladder (n = 4), colon (n = 3), gynecologic (n = 5), and others (n = 3).
RESULTS: One patient had the stent changed after 12 months, and 3 patients had their stents changed at 6 months. These stents were draining adequately with minimal encrustation. Four patients are still alive with functioning stents in situ for 2-10 months. Seven patients died with functioning stents in place (follow-up periods of 1 week to 8 months). Three stents failed from the outset due to bulky pelvic malignancy resulting in high intravesical pressure, as occurs with conventional plastic stents.
CONCLUSION: Our initial experience with the Resonance metallic ureteric stent indicates that it may provide adequate long-term urinary drainage (up to 12 months) in patients with malignant ureteric obstruction but without significantly bulky pelvic disease. This obviates the need for regular stent changes and would offer significant benefit for these patients with limited life expectancy.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17508238     DOI: 10.1007/s00270-007-9043-4

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  14 in total

1.  Update on ureteral stent technology.

Authors:  Dirk Lange; Ben H Chew
Journal:  Ther Adv Urol       Date:  2009-08

Review 2.  Palliative management of malignant upper urinary tract obstruction.

Authors:  P Sountoulides; I Mykoniatis; N Dimasis
Journal:  Hippokratia       Date:  2014 Oct-Dec       Impact factor: 0.471

Review 3.  [Management of ureteral obstruction : Value of percutaneous nephrostomy and ureteral stents].

Authors:  C Netsch; B Becker; A J Gross
Journal:  Urologe A       Date:  2016-11       Impact factor: 0.639

Review 4.  Ureteral stent symptoms and associated infections: a biomaterials perspective.

Authors:  Ben H Chew; Dirk Lange
Journal:  Nat Rev Urol       Date:  2009-07-14       Impact factor: 14.432

5.  Metallic ureteral stents in malignant ureteral obstruction: short-term results and radiological features predicting stent failure in patients with non-urological malignancies.

Authors:  Po-Ming Chow; Jui-Shan Hsu; Shuo-Meng Wang; Hong-Jheng Yu; Yeong-Shiau Pu; Kao-Lang Liu
Journal:  World J Urol       Date:  2013-08-10       Impact factor: 4.226

6.  Clinical experience with ureteral metal stents.

Authors:  Abdulrahman Al Aown; Kyriazis Iason; Kallidonis Panagiotis; Evangelos N Liatsikos
Journal:  Indian J Urol       Date:  2010-10

Review 7.  The full metallic double-pigtail ureteral stent: Review of the clinical outcome and current status.

Authors:  Panagiotis S Kallidonis; Ioannis S Georgiopoulos; Iason D Kyriazis; Stavros Kontogiannis; Abdulrahman M Al-Aown; Evangelos N Liatsikos
Journal:  Indian J Urol       Date:  2015 Jan-Mar

8.  The resonance® metallic ureteric stent in the treatment of chronic ureteric obstruction: a safety and efficacy analysis from a contemporary clinical series.

Authors:  C Patel; D Loughran; R Jones; M Abdulmajed; I Shergill
Journal:  BMC Urol       Date:  2017-03-10       Impact factor: 2.264

9.  Updates on the use of ureteral stents: focus on the Resonance(®) stent.

Authors:  Manoj V Rao; Anthony J Polcari; Thomas Mt Turk
Journal:  Med Devices (Auckl)       Date:  2010-12-22

10.  Malignant Ureteral Obstruction: Functional Duration of Metallic versus Polymeric Ureteral Stents.

Authors:  Po-Ming Chow; I-Ni Chiang; Chia-Yen Chen; Kuo-How Huang; Jui-Shan Hsu; Shuo-Meng Wang; Yuan-Ju Lee; Hong-Jeng Yu; Yeong-Shiau Pu; Chao-Yuan Huang
Journal:  PLoS One       Date:  2015-08-12       Impact factor: 3.240

View more