Development of a guarded liquid ionization chamber for clinical dosimetry.

K J Stewart1, A Elliott, J P Seuntjens.   

Abstract

Liquid ionization chambers are considered superior to air-filled chambers in terms of size, energy dependence and perturbation effects. We constructed and tested a liquid ionization chamber for clinical dosimetry, the GLIC-03, with a sensitive volume of approximately 2 mm3. We also examined two methods to correct for general ion recombination in pulsed photon beams: that of Johansson et al, which modifies Boag's theory for recombination in gases, and an empirical method relating recombination to dose per pulse. The second method can be used even in cases where the first method is not applicable. The response of the GLIC-03 showed a stable, linear and reproducible decrease of 1% over 10 h. The liquid-filled GLIC-03 had a 1.1 +/- 0.4% energy dependence while that of the air-filled GLIC-03 was 2.1 +/- 0.3% between the 6 and 18 MV beams from a Clinac 21EX. The two methods for recombination correction agreed within 0.2% for measurements at 18 MV, 700 V, 100 MU min(-1). Measurements with the GLIC-03 in Solid Water in the build-up region of an 18 MV beam agreed with extrapolation chamber measurements within 1.4%, indicating that the GLIC-03 causes minimal perturbation.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17505091     DOI: 10.1088/0031-9155/52/11/011

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  General ion recombination effect in a liquid ionization chamber in high-dose-rate pulsed photon and electron beams.

Authors:  Hideyuki Takei; Tomonori Isobe; Nozomi Kitamura; Yutaro Mori; Tetsuya Tomita; Daisuke Kobayashi; Satoshi Kamizawa; Tomoharu Sato; Hideyuki Sakurai; Takeji Sakae
Journal:  J Radiat Res       Date:  2018-05-01       Impact factor: 2.724

2.  Pretreatment Dose Verification in Volumetric Modulated Arc Therapy Using Liquid Ionization Chamber.

Authors:  Nitin R Kakade; Rajesh Kumar; Sunil Dutt Sharma; Vikram Mittal; D Datta
Journal:  J Med Phys       Date:  2019 Jan-Mar

3.  Characterization of recombination effects in a liquid ionization chamber used for the dosimetry of a radiosurgical accelerator.

Authors:  Antoine Wagner; Frederik Crop; Thomas Lacornerie; Nick Reynaert
Journal:  J Vis Exp       Date:  2014-05-09       Impact factor: 1.355

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.