[Angioplasty guidewire entrapment affer stent implantation: report of two cases and review of the literature].

Vicens Martí1, Leonel Markarian.   

Abstract

Angioplasty guidewire entrapment, with or without retention of a fractured segment, is an infrequent complication and may lead to acute coronary ischemia due to thromboembolic occlusion. Management can be surgical, conservative or percutaneous. We describe two patients in whom the guidewire remained entrapped with the stent during the pullback maneuver after successful stent implantation. In one patient, we managed to extricate a metallic guidewire using the balloon as a wedge. In the other patient, a fragment of the non-metallic guidewire tip, with hydrophilic coating, remained trapped after being "cut" by the stent during traction towards the ostium while attempting removal. The retained segment was isolated from the blood stream by implanting a synthetic coronary stent graft. We emphasize the high risk of "cutting" this type of non-metallic guidewire in case of engagement with the stent, and we review the literature of this rare complication.

Entities:  

Mesh:

Year:  2007        PMID: 17500193

Source DB:  PubMed          Journal:  Arch Cardiol Mex        ISSN: 1665-1731


  2 in total

1.  Broken guidewire fragment.

Authors:  Kanber Ocal Karabay; Bayram Bagirtan
Journal:  Int J Angiol       Date:  2012-12

2.  Successful retrieval of fractured pressure wire tip (FFR) by hybrid technique.

Authors:  Prakash Sadashivappa Surhonne; Himanshu Mahla; Shivakumar Bhairappa; Shankar Somanna; Cholenahally Nanjappa Manjunath
Journal:  J Saudi Heart Assoc       Date:  2014-09-16
  2 in total