Embolectomy in a rabbit acute arterial occlusion model using a novel electromechanical extraction device.

J Hartman1, W Small, T S Wilson, J Brock, P R Buckley, W J Benett, J M Loge, D J Maitland.   

Abstract

A prototype endovascular electromechanical clot-extraction device was fabricated using a combination of shape memory polymer and shape memory nickel-titanium alloy (nitinol). Five embolic vascular occlusions were created in 4 rabbits by injecting thermally coagulated blood through a 4F catheter in the common carotid artery. Angiography immediately after clot injection showed complete or partial occlusion of the common carotid artery. Posttreatment angiography showed complete (2/5), partial (2/5), or no (1/5) restoration of blood flow.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17494660      PMCID: PMC8134337     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  4 in total

1.  Photothermal properties of shape memory polymer micro-actuators for treating stroke.

Authors:  Duncan J Maitland; Melodie F Metzger; Daniel Schumann; Abraham Lee; Thomas S Wilson
Journal:  Lasers Surg Med       Date:  2002       Impact factor: 4.025

Review 2.  New and future endovascular treatment strategies for acute ischemic stroke.

Authors:  Gary M Nesbit; George Luh; Raymond Tien; Stanley L Barnwell
Journal:  J Vasc Interv Radiol       Date:  2004-01       Impact factor: 3.464

3.  Safety and efficacy of mechanical embolectomy in acute ischemic stroke: results of the MERCI trial.

Authors:  Wade S Smith; Gene Sung; Sidney Starkman; Jeffrey L Saver; Chelsea S Kidwell; Y Pierre Gobin; Helmi L Lutsep; Gary M Nesbit; Thomas Grobelny; Marilyn M Rymer; Isaac E Silverman; Randall T Higashida; Ronald F Budzik; Michael P Marks
Journal:  Stroke       Date:  2005-06-16       Impact factor: 7.914

4.  MERCI 1: a phase 1 study of Mechanical Embolus Removal in Cerebral Ischemia.

Authors:  Y Pierre Gobin; Sidney Starkman; Gary R Duckwiler; Thomas Grobelny; Chelsea S Kidwell; Reza Jahan; John Pile-Spellman; Alan Segal; Fernando Vinuela; Jeffrey L Saver
Journal:  Stroke       Date:  2004-10-28       Impact factor: 7.914

  4 in total
  3 in total

1.  Biomedical applications of thermally activated shape memory polymers.

Authors:  Ward Small; Pooja Singhal; Thomas S Wilson; Duncan J Maitland
Journal:  J Mater Chem       Date:  2010-05-14

2.  Magnetic resonance flow velocity and temperature mapping of a shape memory polymer foam device.

Authors:  Ward Small; Erica Gjersing; Julie L Herberg; Thomas S Wilson; Duncan J Maitland
Journal:  Biomed Eng Online       Date:  2009-12-31       Impact factor: 2.819

3.  Fabrication and in vitro deployment of a laser-activated shape memory polymer vascular stent.

Authors:  Géraldine M Baer; Ward Small; Thomas S Wilson; William J Benett; Dennis L Matthews; Jonathan Hartman; Duncan J Maitland
Journal:  Biomed Eng Online       Date:  2007-11-27       Impact factor: 2.819

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.