Literature DB >> 17487974

Method to reduce chemical background interference in atmospheric pressure ionization liquid chromatography-mass spectrometry using exclusive reactions with the chemical reagent dimethyl disulfide.

Xinghua Guo1, Andries P Bruins, Thomas R Covey.   

Abstract

The interference of chemical background ions (chemical noise) has been a problem since the inception of mass spectrometry. We present here a novel method to reduce the chemical noise in LC-MS based on exclusive gas-phase reactions with a reactive collision gas in a triple-quadrupole mass spectrometer. Combined with the zero neutral loss (ZNL) scan of a triple-quadrupole mass spectrometer, the reactive chemical noise ions can be removed because of shifts of mass-to-charge ratios from the original background ions. The test on various classes of compounds with different functional groups indicates a generic application of this technique in LC-MS. The preliminary results show that a reduction of the level of LC-MS base-peak chromatographic baseline by a factor up to 40 and an improvement of the signal-to-noise ratio by a factor up to 5-10 are achieved on both commercial and custom-modified triple-quadrupole LC-MS systems. Application is foreseen in both quantitative and qualitative trace analysis. It is expected that this chemical noise reduction technique can be optimized on a dedicated mass spectrometric instrumentation which incorporates both a chemical reaction cell for noise reduction and a collision stage for fragmentation.

Entities:  

Year:  2007        PMID: 17487974     DOI: 10.1021/ac062365t

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  2 in total

1.  Biases in ion transmission through an electrospray ionization-mass spectrometry capillary inlet.

Authors:  Jason S Page; Ioan Marginean; Erin S Baker; Ryan T Kelly; Keqi Tang; Richard D Smith
Journal:  J Am Soc Mass Spectrom       Date:  2009-09-02       Impact factor: 3.109

2.  Alternative reagents for chemical noise reduction in liquid chromatography-mass spectrometry using selective ion-molecule reactions.

Authors:  Xinghua Guo; Andries P Bruins; Thomas R Covey; Martin Trötzmüller; Ernst Lankmayr
Journal:  J Am Soc Mass Spectrom       Date:  2008-09-30       Impact factor: 3.109

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.