Evaluation of a portable clinical analyzer in cynomolgus Macaques (Macaca fasicularis).

Kelly J Hopper1, Carolyn Cray.   

Abstract

The aim of this study was to evaluate the clinical performance of a portable analyzer for use in cynomolgus monkeys (Macaca fasicularis). During semiannual health screening, blood samples from 23 animals were analyzed by both the portable clinical analyzer and the institutional comparative pathology laboratory. Portable clinical analyzers have been evaluated for use in other species, but the suitability for macaques has not yet been determined. Results for glucose, urea (BUN), sodium, potassium, chloride, hematocrit, hemoglobin, and total CO2 were compared by overall t test, paired t test, and Pearson correlation. Only glucose and BUN did not differ in the overall t test between methods. Only potassium values did not differ in the paired t test. Compared with those from the portable analyzer, laboratory values were lower for glucose, hematocrit, hemoglobin, and total CO2 and higher for electrolytes and BUN. All values were within normal ranges for cynomolgus macaques which, in this study, were all apparently healthy, physiologically normal animals. We attributed differences between methods to sample type and handling and the physiologic changes in blood after collection. These results indicate that direct comparison of values obtained through different methods may not be valid, and normal ranges for point-of-care devices should be developed for each species.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17487954

Source DB:  PubMed          Journal:  J Am Assoc Lab Anim Sci        ISSN: 1559-6109            Impact factor:   1.232


  3 in total

1.  Evaluation of four hematology and a chemistry portable benchtop analyzers using non-human primate blood.

Authors:  C L Snider; E J Dick; D L McGlasson; M C Robbins; R L Sholund; Y R Bommineni; G B Hubbard
Journal:  J Med Primatol       Date:  2009-09-28       Impact factor: 0.667

2.  Use of portable blood physiology point-of-care devices for basic and applied research on vertebrates: a review.

Authors:  Lauren J Stoot; Nicholas A Cairns; Felicia Cull; Jessica J Taylor; Jennifer D Jeffrey; Félix Morin; John W Mandelman; Timothy D Clark; Steven J Cooke
Journal:  Conserv Physiol       Date:  2014-04-04       Impact factor: 3.079

3.  Validation of the i-STAT and HemoCue systems for the analysis of blood parameters in the bar-headed goose, Anser indicus.

Authors:  T S Harter; M Reichert; C J Brauner; W K Milsom
Journal:  Conserv Physiol       Date:  2015-05-19       Impact factor: 3.079

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.