Literature DB >> 17486872

Electrostatic potential generated during extracorporeal pump prime circulation before cardiopulmonary bypass initiation.

Haley S Newton1, Scott D Niles, James Ploessl, Wayne Richenbacher.   

Abstract

The development of electrostatic potentials generated during cardiopulmonary bypass (CPB) procedures using polyvinylchloride (PVC) tubing in conjunction with roller pumps has been previously documented. The resulting damage from the electrostatic discharge (ESD) has been reported to affect gas transfer devices, but details of potential damage to electronic components commonly used during extracorporeal circulation have not been similarly described. The purpose of this study was to measure the ability of a triboelectric potential to be generated from a primed, circulating, adult CPB pump before the initiation of CPB. Two identical adult CPB circuits were assembled: one incorporating a roller pump and the second incorporating a centrifugal pump mechanism. Primed pumps were circulated (1-6 LPM), and evidence of generated triboelectric potentials was evaluated using a digital multimeter (Fluke 8062 A). The ESD generated from an adult CPB circuit using a roller head configuration elicited a charge in excess of 600 DC V. An identical circuit constructed with a centrifugal pump mechanism did not produce any measurable charge. Sensitive electrical components in the CPB hardware platform may be damaged by ESD potential spikes of this magnitude. Preventative measures, such as circuit charge dissipation, may reduce the potential for such damage when using PVC tubing.

Entities:  

Mesh:

Year:  2007        PMID: 17486872      PMCID: PMC4680680     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  3 in total

1.  Non-occlusive condition with the Better-Header roller pump: impacts of flow dynamics and hemolysis.

Authors:  Eiki Tayama; Hideki Teshima; Tohru Takaseya; Shuji Fukunaga; Kei-ichiro Tayama; Nobuhiko Hayashida; Hidetoshi Akashi; Takemi Kawara; Shigeaki Aoyagi
Journal:  Ann Thorac Cardiovasc Surg       Date:  2004-12       Impact factor: 1.520

2.  Study on causes and prevention of electrostatic charge build-up during extracorporeal circulation.

Authors:  J Snijders; P de Bruijn; M Bergmans; G Bastianen
Journal:  Perfusion       Date:  1999-09       Impact factor: 1.972

3.  Investigation of the phenomenon of electrostatic compromise of a plastic fiber heat exchanger.

Authors:  R J Elgas
Journal:  Perfusion       Date:  1999-03       Impact factor: 1.972