BACKGROUND AND AIM OF THE STUDY: The clinical performance of mechanical heart valves and valve-related complications are important safety endpoints in patients after heart valve replacement. In this retrospective analysis, the mid- to long-term clinical outcomes of two similar bileaflet heart valves, routinely implanted at the authors' institution over an 11-year period, were compared.
METHODS: Between January 1993 and December 2003, a total of 1,161 patients (758 males, 403 females) received either a St. Jude Medical (SJM) or ATS mechanical heart valve. Follow up was obtained via an in-house Quality Management Database, ascertained by telephone questionnaire of the patients and/or family physicians. Follow up was 98.9% complete; the median follow up was 4.6 years; total follow up was 5,624 patient-years (pt-yr).
RESULTS: A total of 604 SJM and 601 ATS prostheses was implanted as isolated (n = 669) or combined (n = 492) procedures. The overall 30-day mortality for SJM and ATS was 4.1% and 3.4%, respectively (p = 0.45). Cumulative survival and freedom from valve-related mortality at 10 years for SJM and ATS valves were 66 +/- 3% versus 68 +/- 5% (p = 0.84) and 96 +/- 1% versus 97 +/- 1% (p = 0.36), respectively. No structural valve failure was encountered for both valve types. Freedom from overall valve-related complications at 10 years was 79 +/- 4% for SJM and 66 +/- 6% for ATS (p = 0.08). The linearized rates for valve-related adverse events for SJM and ATS valves, respectively, were: thromboembolism 0.9 and 1.1%/pt-yr; major bleeding requiring transfusion 0.3 and 0.5%/pt-yr; prosthetic endocarditis 0.03 and 0.1%/pt-yr; paravalvular leak 0.1 and 0.6%/pt-yr.
CONCLUSION: On the basis of an 11-year experience, both bileaflet valves showed very good clinical results, with low incidences of adverse events during the mid-term outcome. Gender and/or concomitant coronary artery disease were not predictors for reduced life expectancy.

Entities:  

Mesh:

Year:  2007        PMID: 17484464

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  5 in total

1.  Device Thrombogenicity Emulator (DTE)--design optimization methodology for cardiovascular devices: a study in two bileaflet MHV designs.

Authors:  Michalis Xenos; Gaurav Girdhar; Yared Alemu; Jolyon Jesty; Marvin Slepian; Shmuel Einav; Danny Bluestein
Journal:  J Biomech       Date:  2010-05-21       Impact factor: 2.712

2.  Percutaneous closure of mitral paravalvular leak via retrograde approach: with use of the Amplatzer duct occluder II and without a wire loop.

Authors:  Hurkan Kursaklioglu; Cem Barcin; Atilla Iyisoy; Oben Baysan; Turgay Celik; Sedat Kose
Journal:  Tex Heart Inst J       Date:  2010

Review 3.  Valve-related complications after mechanical heart valve implantation.

Authors:  Yoshio Misawa
Journal:  Surg Today       Date:  2014-12-18       Impact factor: 2.549

4.  Clinical mid-term outcomes of the Chinese-made CL-V bileaflet mechanical heart valve in Chinese patients.

Authors:  Liang Qi; Shidong Liu; Yan Qiang; Honglin Zhao; Ruisheng Liu; Dianwei Cheng; Bing Song; Quanlin Guan
Journal:  J Thorac Dis       Date:  2021-01       Impact factor: 2.895

5.  Prognosis after surgical replacement with a bioprosthetic aortic valve in patients with severe symptomatic aortic stenosis: systematic review of observational studies.

Authors:  Farid Foroutan; Gordon H Guyatt; Kathleen O'Brien; Eva Bain; Madeleine Stein; Sai Bhagra; Daegan Sit; Rakhshan Kamran; Yaping Chang; Tahira Devji; Hassan Mir; Veena Manja; Toni Schofield; Reed A Siemieniuk; Thomas Agoritsas; Rodrigo Bagur; Catherine M Otto; Per O Vandvik
Journal:  BMJ       Date:  2016-09-28
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.