A mechanical device for aortic compliance modulation: in vitro simulation of aortic dissection treatment.

Romano Zannoli1, Ivan Corazza, Alberto Cremonesi, Angelo Branzi.   

Abstract

Stanford type A aortic dissection often rapidly leads to death from aortic rupture. We considered the possibility of introducing a passive counterpulsating damper into the dissected aorta in order to limit the physical stress associated with ventricular ejection and increase the diastolic aorto-ventricular pressure gradient. We conceived a damping device comprising an intravascular balloon connected to an adjustable external reservoir to regulate the air pressure inside the balloon, and performed a simulation study using a mechanical model of the cardiovascular system, mimicking aortic dissection. When the balloon was completely deflated, the behavior of the aortic pressure signal was typical of low-compliance aortic dissection, as characterized by an augmented maximum systolic value, accentuated diastolic decay, and a very low end-diastolic value. Balloon inflation (at incremental steps to 90 mmHg) progressively restored the aorto-ventricular pressure gradient and reduced peak systolic pressure values, leading to progressive improvements in the characteristics of the aortic pressure curve in terms of reduction in the maximum systolic value and slower diastolic decay. The proposed mechanism might exert beneficial effects at two levels: (1) directly, by reducing mechanical stress on the arterial wall; (2) indirectly, by allowing safer use of pharmaceutical agents (beta-blockers etc.). In vivo animal simulation studies are warranted to verify the effects of the device and optimize balloon shape and volume in a realistic pathophysiological setting.

Entities:  

Mesh:

Year:  2007        PMID: 17482198     DOI: 10.1016/j.jbiomech.2007.03.021

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  2 in total

1.  Wireless Endocardial Atrial (and Ventricular) Sensing with no Implanted Power Source: a Proposal.

Authors:  Ivan Corazza; Igor Diemberger; Christian Martignani; Matteo Ziacchi; Pier Luca Rossi; Alessandro Lombi; Romano Zannoli; Mauro Biffi
Journal:  J Med Syst       Date:  2019-04-26       Impact factor: 4.460

2.  The functional limits of the aneurysmal aortic root. A unique pressure testing apparatus.

Authors:  Timothy Luke Surman; John Matthew Abrahams; Dermot O'Rourke; Karen Jane Reynolds; James Edwards; Michael George Worthington; John Beltrame
Journal:  J Cardiothorac Surg       Date:  2020-09-17       Impact factor: 1.637

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.