Bedside monitoring of subcutaneous interstitial glucose in healthy individuals using microdialysis and infrared spectrometry.

H Michael Heise1, Uwe Damm, Manfred Bodenlenz, Venkata Radhakrishna Kondepati, Gerd Köhler, Martin Ellmerer.   

Abstract

An IR-spectroscopy-based bedside device, coupled to a subcutaneously implanted microdialysis probe, is developed for quasicontinuous glucose monitoring with intermittent readouts at 10-min intervals, avoiding any sensor recalibration under long-term operation. The simultaneous estimation of the microdialysis recovery rate is possible using an acetate containing perfusate and determining its losses across the dialysis membrane. Measurements are carried out on four subjects, with experiments lasting either 8 or 28 h, respectively. Using the spectral interval data either from 1180 to 950 or 1560 to 1000 cm(-1), standard errors of prediction (SEPs) between 0.13 and 0.28 mM are achieved using multivariate calibration with partial least-squares (PLS) or classical least-squares (CLS) calibration models, respectively. The transfer of a PLS calibration model using the spectral and reference concentration data of the dialysates from the three 8-h-long experiments to a 28-h monitoring episode with another healthy subject is tested. Including microdialysis recovery for the determination of the interstitial glucose concentrations, an SEP of 0.24 mM is obtained versus whole blood glucose values. The option to determine other metabolites such as urea or lactate offers the possibility to develop a calibration- and reagent-free point-of-care analyzer.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17477719     DOI: 10.1117/1.2714907

Source DB:  PubMed          Journal:  J Biomed Opt        ISSN: 1083-3668            Impact factor:   3.170


  5 in total

1.  A stepwise approach toward closed-loop blood glucose control for intensive care unit patients: results from a feasibility study in type 1 diabetic subjects using vascular microdialysis with infrared spectrometry and a model predictive control algorithm.

Authors:  Franz Feichtner; Julia K Mader; Roland Schaller; Lukas Schaupp; Martin Ellmerer; Stefan Korsatko; Venkata R Kondepati; H Michael Heise; Malgorzata E Wilinska; Roman Hovorka; Thomas R Pieber
Journal:  J Diabetes Sci Technol       Date:  2011-07-01

Review 2.  Clinical need for continuous glucose monitoring in the hospital.

Authors:  Jeffrey I Joseph; Brian Hipszer; Boris Mraovic; Inna Chervoneva; Mark Joseph; Zvi Grunwald
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

Review 3.  Semi-Implantable Bioelectronics.

Authors:  Jiaru Fang; Shuang Huang; Fanmao Liu; Gen He; Xiangling Li; Xinshuo Huang; Hui-Jiuan Chen; Xi Xie
Journal:  Nanomicro Lett       Date:  2022-05-28

Review 4.  Monitoring Neurochemistry in Traumatic Brain Injury Patients Using Microdialysis Integrated with Biosensors: A Review.

Authors:  Chisomo Zimphango; Farah C Alimagham; Keri L H Carpenter; Peter J Hutchinson; Tanya Hutter
Journal:  Metabolites       Date:  2022-04-26

5.  An automated discontinuous venous blood sampling system for ex vivo glucose determination in humans.

Authors:  Roland Schaller; Franz Feichtner; Hans Köhler; Manfred Bodenlenz; Johannes Plank; Andrea Wutte; Julia K Mader; Martin Ellmerer; Reinhard Hainisch; Thomas R Pieber; Lukas Schaupp
Journal:  J Diabetes Sci Technol       Date:  2009-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.