BACKGROUND: Level of anesthesia may be predicted with the auditory evoked potential or with passive processed electroencephalogram (EEG) parameters. Some previous reports suggest the passive EEG does not reliably predict level of anesthesia in infants. The AAI-1.6 is a relatively new index derived from the AEP/2 monitor. It combines auditory evoked potentials and passive EEG parameters into a single index. This study aimed to assess the AAI-1.6 as a predictor of level of anesthesia in infants and children.
METHODS: Four infants aged less than 1 year, and five older children aged between 2 and 11 years were enrolled. They all had uniform sevoflurane anesthesia for cardiac catheterization. The AAI-1.6 and bispectral index (BIS) were recorded after achieving equilibrium at 1.5%, 2% and 2.5% sevoflurane, and immediately prior to awakening. The prediction coefficient (Pk) for BIS and AAI-1.6 was calculated and compared within each age group.
RESULTS: The Pk for the AAI-1.6 was low in both 0-1 and 2-11 years age groups. In the 2-12 years group, the Pk for BIS was significantly higher than the Pk for the AAI-1.6 (Pk for BIS: 0.89, Pk for AAI-1.6: 0.53, P < 0.01). In contrast in the 0-1 year age group there was no evidence for a difference between the Pk for BIS and the Pk for the AAI-1.6 (Pk for BIS: 0.74, Pk for AAI-1.6: 0.53, P = 0.25).
CONCLUSIONS: This preliminary study suggests AAI-1.6 is a poor predictor of sevoflurane concentration in infants and children.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17474952     DOI: 10.1111/j.1460-9592.2006.02155.x

Source DB:  PubMed          Journal:  Paediatr Anaesth        ISSN: 1155-5645            Impact factor:   2.556


  2 in total

1.  Composite-, plain-auditory evoked potentials index and bispectral index to measure the effects of sevoflurane.

Authors:  Tomoki Nishiyama
Journal:  J Clin Monit Comput       Date:  2013-02-17       Impact factor: 2.502

2.  Correlation between bispectral index, end-tidal anaesthetic gas concentration and difference in inspired-end-tidal oxygen concentration as measures of anaesthetic depth in paediatric patients posted for short surgical procedures.

Authors:  Raylene Dias; Nandini Dave; Barkha Agrawal; Aarti Baghele
Journal:  Indian J Anaesth       Date:  2019-04
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.