| Literature DB >> 17473349 |
Nina Tilly1, Erik Grusell, Peter Kimstrand, Stefan Lorin, Konrad Gajewski, Mikael Pettersson, Andreas Bäcklund, Bengt Glimelius.
Abstract
In this paper we present the recent developments made for the scanning system for proton beams at TSL in Uppsala, showing that this system is now fully functional being able to produce conformal intensity modulated scan patterns with sufficient accuracy. A new control and supervising system handling the beam delivery including the control of the synchrocyclotron and the scanning system is developed and described in detail. A complete dosimetry system with transmission ionization chambers and a multi-wire ionization chamber for monitoring of the beam during scanning has been constructed. The details of the dose monitors and the position sensitive multi-wire ionization chamber are presented in this work. Furthermore, we have established procedures for verification measurements to ensure the quality of the beam and also methods for calibration of the beam monitors and relative and absolute dosimetry for complex scanned beams.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17473349 DOI: 10.1088/0031-9155/52/10/008
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609