Literature DB >> 17463125

The effect of tourniquet padding on the efficiency of tourniquets of the upper limb.

A Rajpura1, B V Somanchi, L T S W Muir.   

Abstract

We report the effect of padding on the efficiency of the pneumatic tourniquet for the upper limb. Varying thicknesses of two commercially-available types of orthopaedic padding (Cellona and Velband) were applied to the arms of 20 volunteers, with three pressure transducers placed directly beneath the padding. A tourniquet was positioned over the padding and inflated to 220 mmHg. Significant reductions in the transmitted pressure were recorded from the transducers with both padding materials. With eight layers of padding, reductions in pressure of 13% (1% to 26%) and 18% (7% to 35%) were seen with Cellona and Velband, respectively. The reduction in pressure with Velband padding correlated with increasing arm circumference (Pearson's correlation coefficient 0.711, p < 0.001). Studies to date have examined how arm circumference affects the required tourniquet inflation pressure. Our study is the first to investigate the effect of the padding and the findings suggest that using more than two layers results in a significant reduction in the transmitted pressure.

Mesh:

Year:  2007        PMID: 17463125     DOI: 10.1302/0301-620X.89B4.18778

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  3 in total

1.  Safety and efficacy of a new tourniquet system.

Authors:  Junko Sato; Yoshinori Ishii; Hideo Noguchi; Mitsuhiro Takeda
Journal:  BMC Surg       Date:  2012-08-15       Impact factor: 2.102

Review 2.  Tourniquet in surgery of the limbs: a review of history, types and complications.

Authors:  Alireza Saied; Alia Ayatollahi Mousavi; Fateme Arabnejad; Afshin Ahmadzadeh Heshmati
Journal:  Iran Red Crescent Med J       Date:  2015-02-18       Impact factor: 0.611

3.  In-vivo analysis of epicutaneous pressure distribution beneath a femoral tourniquet--an observational study.

Authors:  Klaus Edgar Roth; Boris Mandryka; Gerrit Steffen Maier; Uwe Maus; Manfred Berres; Jan-Dirk Rompe; Friedrich Bodem
Journal:  BMC Musculoskelet Disord       Date:  2015-01-31       Impact factor: 2.362

  3 in total