Validation of a quality assessment system for blood gas and electrolyte testing.

John G Toffaletti1, Elizabeth H McDonnell, Lakshmi V Ramanathan, Judit Tolnai, Rozanna Templin, Lynn Pompa.   

Abstract

BACKGROUND: A recently-introduced quality assessment system (Intelligent Quality Management: iQM), was evaluated in routine clinical use at four different hospitals. The iQM technology is designed to replace conventional external liquid controls with software, Process Control (PC) Solutions and Calibration Validation components that continually assess the function of the GEM Premier 3000 (GEM) analyzer and automatically initiate and document corrective actions.
METHODS: We validated the performance claims of iQM by monitoring quality control (QC) materials at 4 clinical sites while analyzing approximately 10,550 patient samples. We compared iQM-measured QC values to traditional QC results, evaluating the number and type of error flags for patient samples, and used data from control results to calculate the average time to detect an error (ADT) for each analyte.
RESULTS: The calculated ADT was approximately 3 min for all analytes except for sodium (17 min), glucose (11 min), and lactate (5.9 min). Precision of control materials in iQM cartridges was better than from external controls run on traditional analyzers. The iQM system detected errors in 0.46% of actual clinical samples.
CONCLUSIONS: The findings from our study confirm that (a) iQM precision in a clinical setting is comparable to that found in previous studies done in a research setting, (b) the improved precision of control material on the iQM is likely because the internal control fluids are sealed and not susceptible to exposure from handling, and (c) the system detects and often corrects errors in specific samples that might not be reported by traditional analytical systems.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17462612     DOI: 10.1016/j.cca.2007.03.021

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  3 in total

1.  Three-year monitoring and comparison of results from two identical blood gas analyzers.

Authors:  Yun Huang; Robert Dean; Yvonne Dubbelman; Anne Vincent; Ying-Pui Michael Chan
Journal:  Pract Lab Med       Date:  2022-06-03

2.  Intelligent Quality Management 2 with IntraSpect™ technology for quality control of GEM® Premier™ 5000 blood gas analyzers- A novel application of the patient sample as its own control.

Authors:  James O Westgard; Jose Cervera
Journal:  Pract Lab Med       Date:  2022-04-08

3.  Method validation of a set of 12 GEM® Premier™ 4000 blood gas analyzers for point-of-care testing in a university teaching hospital.

Authors:  Charlotte Oris; Yoan Clavel; Matthieu Jabaudon; Annick Pialat; Hadj Abdelkader Mohamed; Frédérique Lioret; Vincent Sapin; Damien Bouvier
Journal:  Pract Lab Med       Date:  2017-12-13
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.