BACKGROUND: This study was undertaken because of the accidental observation that a sample of 60+ beads was obtained by the instrument from a completely dry, unused well in a 96 well plate. Others have observed unexplained outliers in replicated wells. The problem was first observed on an older instrument, and replicated on a new instrument. METHODS ANDEntities:
Mesh:
Substances:
Year: 2007 PMID: 17459164 PMCID: PMC1865554 DOI: 10.1186/1477-5751-6-6
Source DB: PubMed Journal: J Negat Results Biomed ISSN: 1477-5751
Assays and bead classifiers available for use
| 106 | 105 | 104 | |
| 112 | 111 | 147 | 133 |
| 113 | 115 | 152 | 137 |
| 180 | 118 | 197 | |
| 198 | |||
| 173 |
CMV = Cytomegalovirus
SFV = Simian Foamy Virus
SRV = Simian Type D Retrovirus
SIV = Simian Immunodeficiency Virus
Items in bold are duplicated bead identifiers which were not used.
| 3 | 5 | 6 | 9 | 6 | 5 | 4 | 0 | 5 | 6 | 4 | 6 | |
| 12 | 1 | 1 | 7 | 0 | 6 | 0 | 29 | 1 | 3 | 3 | 8 | |
| 6 | 6 | 2 | 1 | 0 | 9 | 0 | 1 | 24 | 14 | 31 | 15 | |
| 3 | 8 | 1 | 5 | 2 | 4 | 2 | 1 | 6 | 1 | 6 | 4 | |
| 1 | 5 | 5 | 2 | 9 | 1 | 0 | 39 | 57 | 36 | 37 | 52 | |
| 23 | 9 | 1 | 0 | 8 | 8 | 114 | 79 | 1 | 62 | 1 | 22 |
(Note, the bead number above was 25. Intra-lab assays have a preceding digit "1", hence 125.)
Data in bold italics is the row with actual sample. All other rows were empty, filled with buffer just prior to reading.
Figure 1Summary of 6 bead sets, trailing empty well contents for instrument B. Bead counts.
Figure 2Summary of 6 bead sets, trailing empty well contents for instrument A. Bead counts.