Mathew R Guilfoyle1, H Fernandes, S Price.
Abstract
We report the unexpected adjustment of a Strata shunt valve by a vagus nerve stimulator-activating magnet, resulting in symptomatic under-drainage of CSF. Implantable devices using magnets for calibration and activation are becoming more prevalent within neurosurgery and the present case highlights the potential for adverse interactions.Entities:
Mesh:
Year: 2007 PMID: 17453775 DOI: 10.1080/02688690701243217