| Literature DB >> 17442639 |
V Tavares1, C C Macedo, L Montanhez, F A P Barros, E C Meurer, D R Campos, E C Coelho, S A Calaffati, J Pedrazzoli.
Abstract
Here we present a sensitive and specific liquid chromatography-tandem mass spectrometric method for the quantification of dimenhydrinate (I) in human plasma. Sample preparation is conducted using citalopram (II) addition as an internal standard (IS), liquid-liquid extraction with basified plasma using a mixture hexane/acetate (1:1, v/v) as the extracting solvent, and the final extract reconstituted in the mobile phase. I and II (IS) were injected in a C8 column with the mobile phase composed of methanol:isopropanol:water:formic acid (78.00:19.92:2.00:0.08, v/v/v/v) and monitored using a positive electrospray source with tandem mass spectrometry analyses. The selected reaction monitoring (SRM) was set using precursor ion and product ion combinations of m/z 256.0>167.0 and m/z 325.0>109.0 for I and II, respectively. The limit of quantification (LOQ) was 0.4 ng/mL, the dynamic range being 0.4-200 ng/mL. Validation results on linearity, specificity, accuracy, precision and stability, as well as on application to the analysis of plasma samples taken up to 24 h after oral administration of 100 mg of dimenhydrinate in healthy volunteers demonstrated its applicability to bioavailability studies.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17442639 DOI: 10.1016/j.jchromb.2007.03.027
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205