In this communication, a theoretical framework for quality control and parameter optimization in diffusion tensor imaging (DTI) is presented and validated. The approach is based on the analytical error propagation of the mean diffusivity (D(av)) obtained directly from the diffusion-weighted data acquired using rotationally invariant and uniformly distributed icosahedral encoding schemes. The error propagation of a recently described and validated cylindrical tensor model is further extrapolated to the spherical tensor case (diffusion anisotropy approximately 0) to relate analytically the precision error in fractional tensor anisotropy (FA) with the mean diffusion-to-noise ratio (DNR). The approach provided simple analytical and empirical quality control measures for optimization of diffusion parameter space in an isotropic medium that can be tested using widely available water phantoms.Mesh:
Year: 2007 PMID: 17442523 PMCID: PMC2185542 DOI: 10.1016/j.mri.2007.02.011
Source DB: PubMed Journal: Magn Reson Imaging ISSN: 0730-725X Impact factor: 2.546