Literature DB >> 17442293

Effect of HbS in the determination of HbA2 with the Biorad Variant II analyzer.

Christoforos Kalleas1, Ioannis Tentes, Dimitrios Margaritis, Kostas Anagnostopoulos, Athina Toli, Dimitrios Pendilas, Georgios Bourikas, Constantinos Tsatalas, Alexandros H Kortsaris.   

Abstract

OBJECTIVES: The effect of the presence of HbS in the determination of HbA2 using the Biorad Variant II analyzer. DESIGN AND METHODS: The effect of HbS presence in the samples was quantified using the HELENA SAS-MX alkaline gel electrophoresis kit as the reference method.
RESULTS: The %HbA2 values from the Variant II analyzer and the HELENA SAS-MX alkaline gel electrophoresis kit show a good linear correlation in the absence of HbS. A strong positive bias in the %HbA2 values from the Variant II is apparent in the presence of HbS in the samples, when compared to the alkaline electrophoresis gel.
CONCLUSION: The Variant II analyzer gives reliable results for %HbA2 determination when no HbS is detectable in the samples. When HbS is present, the gel electrophoresis method gives more accurate results.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17442293     DOI: 10.1016/j.clinbiochem.2007.03.008

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  1 in total

1.  Hemoglobin A2 values in sickle cell disease patients quantified by high performance liquid chromatography and the influence of alpha thalassemia.

Authors:  Silvana Fahel da Fonseca; Tatiana Amorim; Antônio Purificação; Marilda Gonçalves; Ney Boa-Sorte
Journal:  Rev Bras Hematol Hemoter       Date:  2015-06-07
  1 in total