A novel phase assignment protocol and driving system for a high-density focused ultrasound array.

R Erich Caulfield1, Xiangtao Yin, Jose Juste, Kullervo Hynynen.   

Abstract

Currently, most phased-array systems intended for therapy are one-dimensional (1-D) and use between 5 and 200 elements, with a few two-dimensional (2-D) systems using several hundred elements. The move toward lambda/2 interelement spacing, which provides complete 3-D beam steering, would require a large number of closely spaced elements (0.15 mm to 3 mm). A solution to the resulting problem of cost and cable assembly size, which this study examines, is to quantize the phases available at the array input. By connecting elements with similar phases to a single wire, a significant reduction in the number of incoming lines can be achieved while maintaining focusing and beam steering capability. This study has explored the feasibility of such an approach using computer simulations and experiments with a test circuit driving a 100-element linear array. Simulation results demonstrated that adequate focusing can be obtained with only four phase signals without large increases in the grating lobes or the dimensions of the focus. Experiments showed that the method can be implemented in practice, and adequate focusing can be achieved with four phase signals with a reduction of 20% in the peak pressure amplitude squared when compared with the infinite-phase resolution case. Results indicate that the use of this technique would make it possible to drive more than 10,000 elements with 33 input lines. The implementation of this method could have a large impact on ultrasound therapy and diagnostic devices.

Mesh:

Year:  2007        PMID: 17441588     DOI: 10.1109/tuffc.2007.312

Source DB:  PubMed          Journal:  IEEE Trans Ultrason Ferroelectr Freq Control        ISSN: 0885-3010            Impact factor:   2.725


  4 in total

1.  A study of heating duration and scanning path in focused ultrasound surgery.

Authors:  Dehui Li; Guofeng Shen; Hui Luo; Jingfeng Bai; Jinfeng Bai; Yazhu Chen
Journal:  J Med Syst       Date:  2011-10       Impact factor: 4.460

Review 2.  Image-guided ultrasound phased arrays are a disruptive technology for non-invasive therapy.

Authors:  Kullervo Hynynen; Ryan M Jones
Journal:  Phys Med Biol       Date:  2016-08-05       Impact factor: 3.609

Review 3.  Driving circuitry for focused ultrasound noninvasive surgery and drug delivery applications.

Authors:  Munir M El-Desouki; Kullervo Hynynen
Journal:  Sensors (Basel)       Date:  2011-01-07       Impact factor: 3.576

4.  Measurement and simulation of steered acoustic fields generated by a multielement array for therapeutic ultrasound.

Authors:  Eleanor Martin; Morgan Roberts; Bradley Treeby
Journal:  JASA Express Lett       Date:  2021-01-11
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.