Tidal volume delivery during high-frequency oscillatory ventilation in adults with acute respiratory distress syndrome.

David N Hager1, Henry E Fessler, David W Kaczka, Carl B Shanholtz, Mathew K Fuld, Brett A Simon, Roy G Brower.   

Abstract

OBJECTIVE: a) Characterize how ventilator and patient variables affect tidal volume during high-frequency oscillatory ventilation; and b) measure tidal volumes in adults with acute respiratory distress syndrome during high-frequency oscillatory ventilation.
DESIGN: Observational study.
SETTING: Research laboratory and medical intensive care unit. PATIENTS: Test lung and patients with acute respiratory distress syndrome.
INTERVENTIONS: Using a previously validated hot wire anemometer placed in series with a Sensormedics 3100B high-frequency ventilator, an endotracheal tube, and a test lung, tidal volume was measured at different combinations of frequency (4, 6, 8, 10, and 12 Hz), pressure amplitude (50, 60, 70, 80, and 90 cm H2O), mean airway pressure (20, 30, and 40 cm H2O), test lung compliance (10, 30, and 50 mL/cm H2O), endotracheal tube internal diameter (6, 7, and 8 mm), bias flow (20, 30, and 40 L/min), and inspiratory/expiratory ratio (1:2 and 1:1). In patients, tidal volume was measured at baseline ventilator settings and at baseline frequency +/-2 Hz and baseline pressure amplitude +/-10 cm H2O.
MEASUREMENTS AND MAIN RESULTS: Measured tidal volumes were 23-225 mL during high-frequency oscillatory ventilation of the test lung. A 2-Hz increase in frequency and a 10-cm H2O increase in pressure amplitude caused a 21.3% +/- 4.1% decrease and 21.4% +/- 3.4% increase in tidal volume, respectively. Decreasing endotracheal tube internal diameter from 8 mm to 7 mm and from 7 mm to 6 mm caused a 15.3% +/- 1.7% and 18.9% +/- 2.1% reduction in tidal volume, respectively. Increasing bias flow from 20 L/min to 30 L/min increased tidal volume by 11.2% +/- 3.9%. Further increases in bias flow, changes in compliance, and changes in mean airway pressure had little effect. Tidal volumes measured in acute respiratory distress syndrome patients were 44-210 mL. A 2-Hz increase in frequency was associated with a 23.1% +/- 6.3% decrease in tidal volume. In contrast to the test lung data, a 10-cm H2O increase in pressure amplitude resulted in only a 5.6% +/- 4.5% increase in tidal volume.
CONCLUSIONS: Tidal volumes are not uniformly small during high-frequency oscillatory ventilation. The primary determinant of tidal volume in adults with acute respiratory distress syndrome during high-frequency oscillatory ventilation with the Sensormedics 3100B is frequency. Test lung findings suggest that endotracheal tube internal diameter is also an important determinant of tidal volume.

Entities:  

Mesh:

Year:  2007        PMID: 17440422     DOI: 10.1097/01.CCM.0000266586.04676.55

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  20 in total

Review 1.  The design of future pediatric mechanical ventilation trials for acute lung injury.

Authors:  Robinder G Khemani; Christopher J L Newth
Journal:  Am J Respir Crit Care Med       Date:  2010-08-23       Impact factor: 21.405

2.  Airway pressure and transpulmonary pressure during high-frequency oscillation for acute respiratory distress syndrome.

Authors:  William R Henderson; Paolo B Dominelli; Donald E G Griesdale; Daniel Talmor; A William Sheel
Journal:  Can Respir J       Date:  2013-10-17       Impact factor: 2.409

Review 3.  High-frequency oscillation as a rescue strategy for brain-injured adult patients with acute lung injury and acute respiratory distress syndrome.

Authors:  Neil H Young; Peter J D Andrews
Journal:  Neurocrit Care       Date:  2011-12       Impact factor: 3.210

4.  Comparison of high-frequency oscillation and tracheal gas insufflation versus standard high-frequency oscillation at two levels of tracheal pressure.

Authors:  Spyros D Mentzelopoulos; Sotiris Malachias; Stelios Kokkoris; Charis Roussos; Spyros G Zakynthinos
Journal:  Intensive Care Med       Date:  2010-03-16       Impact factor: 17.440

Review 5.  Fifty Years of Research in ARDS. Vt Selection in Acute Respiratory Distress Syndrome.

Authors:  Sarina K Sahetya; Jordi Mancebo; Roy G Brower
Journal:  Am J Respir Crit Care Med       Date:  2017-12-15       Impact factor: 21.405

6.  Volumetric xenon-CT imaging of conventional and high-frequency oscillatory ventilation.

Authors:  Daniel G Mulreany; Brett A Simon; Kieran J Murphy; R Blaine Easley
Journal:  Acad Radiol       Date:  2009-03-05       Impact factor: 3.173

7.  Ventilation in chest trauma.

Authors:  Torsten Richter; Maximilian Ragaller
Journal:  J Emerg Trauma Shock       Date:  2011-04

Review 8.  High-frequency oscillatory ventilation versus conventional ventilation for acute respiratory distress syndrome.

Authors:  Sachin Sud; Maneesh Sud; Jan O Friedrich; Hannah Wunsch; Maureen O Meade; Niall D Ferguson; Neill K J Adhikari
Journal:  Cochrane Database Syst Rev       Date:  2016-04-04

9.  Functional lung imaging during HFV in preterm rabbits.

Authors:  Jordan Thurgood; Stuart Hooper; Melissa Siew; Megan Wallace; Stephen Dubsky; Marcus Kitchen; R Aidan Jamison; Richard Carnibella; Andreas Fouras
Journal:  PLoS One       Date:  2012-10-30       Impact factor: 3.240

10.  Arteriovenous Extracorporeal Lung Assist Allows For Maximization Of Oscillatory Frequencies: A Large-animal Model Of Respiratory Distress.

Authors:  Ralf M Muellenbach; Julian Kuestermann; Markus Kredel; Amélie Johannes; Ulrike Wolfsteiner; Frank Schuster; Christian Wunder; Peter Kranke; Norbert Roewer; Jörg Brederlau
Journal:  BMC Anesthesiol       Date:  2008-11-14       Impact factor: 2.217

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.