| Literature DB >> 17437866 |
Velayutham Gopikrishna1, Kush Tinagupta, Deivanayagam Kandaswamy.
Abstract
Pulse oximeter is a well-established and noninvasive method for measuring vascular health by evaluating oxygen saturation. This study compared the efficacy of a custom-made pulse oximeter dental probe with the electric pulp testing and thermal testing for measuring pulp vitality status of recently traumatized permanent teeth. Readings for pulp vitality for 17 recently traumatized maxillary incisors were taken with custom-made pulse oximeter dental probe (group 1), electrical pulp tester (group 2), and thermal testing (group 3) over a 6-month period. The proportion of recently traumatized teeth showing a positive responsiveness in thermal/electric pulp tests increased from no teeth showing responsiveness on day 0 to 29.4% teeth on the 28th day, 82.35% of teeth at 2 months, and 94.11% teeth at 3 months. However, pulse oximeter gave positive vitality readings that remained constant over the study period from day 0 to 6 months in all patients.Entities:
Mesh:
Year: 2007 PMID: 17437866 DOI: 10.1016/j.joen.2007.01.014
Source DB: PubMed Journal: J Endod ISSN: 0099-2399 Impact factor: 4.171