Literature DB >> 1743189

Radiology of the AS 800 artificial urinary sphincter; normal appearances and complications.

S Hamilton1, H D Flood, M K Shetty, R Grainger.   

Abstract

Thirteen adult male patients who had an AS 800 artificial urinary sphincter inserted are reviewed. Five have had malfunction, and in four the causes were diagnosed radiologically. These included: fistula formation, leakage of hydraulic fluid, air in the system, and inadequate deflation of the cuff. Since the sphincter is filled with contrast medium, it is ideally suited to radiological assessment.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1743189     DOI: 10.1016/0720-048x(91)90093-b

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  1 in total

Review 1.  A complication of Artificial Urinary Sphincter: migration of pump onto the rectus sheath in a patient with cauda equina syndrome.

Authors:  Michelle Christodoulidou; Mathew Thomas; Gurpreet Singh
Journal:  BMJ Case Rep       Date:  2013-07-26
  1 in total