Can oncology patients' central venous catheters be used for isotope assessment of glomerular filtration rate? An in-vitro study.

Beth L Gawthorpe1, Mark C Barnfield, Maria T Burniston.   

Abstract

BACKGROUND: Isotope assessment of glomerular filtration rate (GFR) is frequently performed in patients with central venous catheters (CVCs). Use of the CVCs for administration of tracer and subsequent blood sampling would be less distressing for patients (particularly paediatric) and would reduce the frequency of failed samples due to poor venous access. However, the GFR test is quantitative and could be affected by incomplete tracer delivery due to adhesion to the CVC and also by contamination of blood samples due to adhered tracer leaching back into the sampled blood as it passes through the CVC. AIM: This in vitro study aimed to quantify the effects on GFR assessment of tracer adhesion and leaching, in single-lumen and dual-lumen CVCs.
METHOD: New and clinically used single-lumen CVCs were injected with tracer (99mTc-DTPA and 51Cr-EDTA) and then flushed repeatedly with saline. The outflows were assayed in a gamma counter and, where possible, the CVCs were imaged on a gamma camera to take snap shots of tracer movement throughout a GFR assessment. In a separate experiment, a phantom patient was used to compare blood sampling through a dual lumen CVC with peripheral sampling. RESULT AND
CONCLUSION: A CVC successfully delivers >99% of tracer. Subsequent blood samples can be taken through the other lumen of a dual-lumen CVC but not through a single-lumen as this significantly alters the GFR result due to contamination.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17414892     DOI: 10.1097/MNM.0b013e328082545e

Source DB:  PubMed          Journal:  Nucl Med Commun        ISSN: 0143-3636            Impact factor:   1.690


  2 in total

1.  Focal retention of Tc-99m sestamibi at central venous catheter tip: a potential source of artifact.

Authors:  Charles J Hardebeck
Journal:  J Nucl Cardiol       Date:  2014-02       Impact factor: 5.952

2.  Accurate and precise plasma clearance measurement using four 99mTc-DTPA plasma samples over 4 h.

Authors:  Surajith N Wanasundara; Michal J Wesolowski; Mark C Barnfield; Michael L Waller; Anthony W Murray; Maria T Burniston; Paul S Babyn; Carl A Wesolowski
Journal:  Nucl Med Commun       Date:  2016-01       Impact factor: 1.690

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.