Cortical screw pullout strength and effective shear stress in synthetic third generation composite femurs.

Radovan Zdero1, Shaun Rose, Emil H Schemitsch, Marcello Papini.   

Abstract

BACKGROUND: The use of artificial bone analogs in biomechanical testing of orthopaedic fracture fixation devices has increased, particularly due to the recent development of commercially available femurs such as the third generation composite femur that closely reproduce the bulk mechanical behavior of human cadaveric and/or fresh whole bone. The purpose of this investigation was to measure bone screw pullout forces in composite femurs and determine whether results are comparable to cadaver data from previous literature. METHOD OF APPROACH: The pullout strengths of 3.5 and 4.5 mm standard bicortical screws inserted into synthetic third generation composite femurs were measured and compared to existing adult human cadaveric and animal data from the literature.
RESULTS: For 3.5 mm screws, the measured extraction shear stress in synthetic femurs (23.70-33.99 MPa) was in the range of adult human femurs and tibias (24.4-38.8 MPa). For 4.5 mm screws, the measured values in synthetic femurs (26.04-34.76 MPa) were also similar to adult human specimens (15.9-38.9 MPa). Synthetic femur results for extraction stress showed no statistically significant site-to-site effect for 3.5 and 4.5 mm screws, with one exception. Overall, the 4.5 mm screws showed statistically higher stress required for extraction than 3.5 mm screws.
CONCLUSIONS: The third generation composite femurs provide a satisfactory biomechanical analog to human long-bones at the screw-bone interface. However, it is not known whether these femurs perform similarly to human bone during physiological screw "toggling."

Entities:  

Mesh:

Year:  2007        PMID: 17408335     DOI: 10.1115/1.2540926

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  3 in total

1.  Feasibility and fidelity of practising surgical fixation on a virtual ulna bone.

Authors:  Justin LeBlanc; Carol Hutchison; Yaoping Hu; Tyrone Donnon
Journal:  Can J Surg       Date:  2013-08       Impact factor: 2.089

2.  Comparison and prediction of pullout strength of conical and cylindrical pedicle screws within synthetic bone.

Authors:  Wen-Chi Tsai; Po-Quang Chen; Tung-Wu Lu; Shing-Sheng Wu; Kao-Shang Shih; Shang-Chih Lin
Journal:  BMC Musculoskelet Disord       Date:  2009-04-30       Impact factor: 2.362

3.  Biomechanical analysis of operative methods in the treatment of extra-articular fracture of the proximal tibia.

Authors:  Seong-Man Lee; Chang-Wug Oh; Jong-Keon Oh; Joon-Woo Kim; Hyun-Joo Lee; Chang-Soo Chon; Byoung-Joo Lee; Hee-Soo Kyung
Journal:  Clin Orthop Surg       Date:  2014-08-05
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.