J Taki1, A Muramori, K Nakajima, H Bunko, M Kawasuji, N Tonami, K Hisada.
Abstract
A continuous ventricular function monitor with a miniature cadmium telluride detector was evaluated and applied to patients with coronary bypass surgery (CABG). Ejection fraction (EF) at rest and change in EF from rest to exercise and postexercise (Y) measured with the device correlated with that of the gamma camera (X) (Y = 0.86x + 6.8(%), r = 0.87, n = 110, p less than 0.001, and Y = 0.96x + 0.4(%), r = 0.90, n = 37, p less than 0.001 respectively). Left ventricular function during and after supine ergometer exercise was monitored in 54 patients before and after CABG. The EF change from baseline to peak exercise improved from -5.9% +/- 8.9% before CABG to 7.2% +/- 7.9% after CABG (p less than 0.001). In all patients but two, a rapid EF increase just after exercise over baseline EF was observed. This EF "overshoot" during recovery increased from 11.5% +/- 6.5% to 16.4% +/- 6.0% (p less than 0.001) after CABG. The time from the cessation of exercise to EF overshoot decreased from 153 +/- 80 sec to 76 +/- 49 sec (p less than 0.001) after CABG. The continuous ventricular function monitor with a miniature cadmium telluride detector is able to measure EF reliably. Following successful aortocoronary bypass, EF response during exercise improved and the EF overshoot in the recovery phase became faster and higher.Entities:
Mesh:
Substances:
Year: 1992 PMID: 1740717
Source DB: PubMed Journal: J Nucl Med ISSN: 0161-5505 Impact factor: 10.057