Literature DB >> 17392844

Comparison of silicone and polypropylene Ahmed glaucoma valves: two-year follow-up.

Paul J Mackenzie1, Robert M Schertzer, Carolyn M Isbister.   

Abstract

BACKGROUND: To compare the safety and efficacy of silicone and polypropylene Ahmed glaucoma drainage devices.
METHODS: We studied a retrospective consecutive case series of 91 eyes of 81 patients who were implanted with Ahmed valves by a single surgeon (R.M.S.). The first 43 eyes received polypropylene (model S2) valves; the remaining 48 eyes received silicone (model FP7) valves. Data collection included demographics, type of glaucoma, and sequential measurement of clinical characteristics. The mean outcome measures were visual acuity, intraocular pressure (IOP), number of antiglaucoma medications, and postoperative complications. Data were collected on day 1; weeks 1, 2-3, and 6; months 3-4, 6, and 12; and last visit.
RESULTS: Mean follow-up was 27.1 months for the polypropylene valve and 20.1 months for the silicone valve. Mean decrease in IOP at 1 year was significantly greater in the silicone group than in the polypropylene group (t test, p < 0.005); however, this difference disappeared at 2 years. There were no differences between the groups in visual acuity, number of antiglaucoma medications, bleb needlings with 5-fluorouracil, or minor complications. A greater number of serious complications, including valve failures, occurred in the polypropylene group than in the silicone group, and this difference could not be accounted for by the longer follow-up in the polypropylene group. At 2 years follow-up, 82% of patients with silicone valves and 72% of patients with polypropylene valves continued to achieve greater than 30% IOP reduction.
INTERPRETATION: Both the silicone and polypropylene Ahmed valves significantly reduced IOP over 2 years. The silicone valve was associated with fewer serious complications.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17392844     DOI: 10.3129/can j ophthalmol.i07-032

Source DB:  PubMed          Journal:  Can J Ophthalmol        ISSN: 0008-4182            Impact factor:   1.882


  13 in total

1.  Intraoperative testing of opening and closing pressure predicts risk of low intraocular pressure after Ahmed glaucoma valve implantation.

Authors:  F Bochmann; A Kipfer; J Tarantino; C Kaufmann; L Bachmann; M Thiel
Journal:  Eye (Lond)       Date:  2014-07-25       Impact factor: 3.775

2.  The Ahmed glaucoma valve in neovascular glaucoma (An AOS Thesis).

Authors:  Peter A Netland
Journal:  Trans Am Ophthalmol Soc       Date:  2009-12

3.  Evaluation of Ahmed glaucoma valve implantation through a needle-generated scleral tunnel in Mexican children with glaucoma.

Authors:  Oscar Albis-Donado; Félix Gil-Carrasco; Rafael Romero-Quijada; Ravi Thomas
Journal:  Indian J Ophthalmol       Date:  2010 Sep-Oct       Impact factor: 1.848

4.  Inhibition of biomaterial-induced complement activation attenuates the inflammatory host response to implantation.

Authors:  Ioannis Kourtzelis; Stavros Rafail; Robert A DeAngelis; Periklis G Foukas; Daniel Ricklin; John D Lambris
Journal:  FASEB J       Date:  2013-04-04       Impact factor: 5.191

5.  Polypropylene vs silicone Ahmed valve with adjunctive mitomycin C in paediatric age group: a prospective controlled study.

Authors:  Y El Sayed; A Awadein
Journal:  Eye (Lond)       Date:  2013-04-12       Impact factor: 3.775

6.  Development of an Experimental Drug Eluting Suprachoroidal Microstent as Glaucoma Drainage Device.

Authors:  Marina Hovakimyan; Stefan Siewert; Wolfram Schmidt; Katrin Sternberg; Thomas Reske; Oliver Stachs; Rudolf Guthoff; Andreas Wree; Martin Witt; Klaus-Peter Schmitz; Reto Allemann
Journal:  Transl Vis Sci Technol       Date:  2015-06-30       Impact factor: 3.283

7.  Risk factors for hypertensive phase after Ahmed glaucoma valve implantation.

Authors:  Onur Özalp; Serdar İlgüy; Eray Atalay; Tülay Şimşek; Nilgün Yıldırım
Journal:  Int Ophthalmol       Date:  2021-08-22       Impact factor: 2.031

8.  Comparison of the Ahmed glaucoma valve with the Baerveldt glaucoma implant: a meta-analysis.

Authors:  Yi-Wen Wang; Ping-Bao Wang; Chao Zeng; Xiao-Bo Xia
Journal:  BMC Ophthalmol       Date:  2015-10-13       Impact factor: 2.209

Review 9.  Ahmed glaucoma valve implant: surgical technique and complications.

Authors:  Ivano Riva; Gloria Roberti; Francesco Oddone; Anastasios Gp Konstas; Luciano Quaranta
Journal:  Clin Ophthalmol       Date:  2017-02-17

10.  Clinical Experience with the M4 Ahmed Glaucoma Drainage Implant.

Authors:  Ilya Sluch; Brett Gudgel; Justin Dvorak; Mary Anne Ahluwalia; Kai Ding; Steve Vold; Steven Sarkisian
Journal:  J Curr Glaucoma Pract       Date:  2017-10-27
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.