Water permeability of high-flux dialyzer membranes after Renalin reprocessing.

M E Labib1, J Murawski, Y Tabani, S H Wolff, A L Zydney, F R Funderburk, Z Huang, T Kapoian, R A Sherman.   

Abstract

Dialysis with high-flux membranes is widely used, in part, because they are thought to increase the removal of middle molecules when compared with low-flux membranes. Dialyzer reprocessing; however, is thought to alter middle molecule clearance. Renalin, a mixture of germicidal agents, has widespread use in dialyzer reprocessing. We determined the effect of Renalin reprocessing on the water permeability of three different dialyzers of Fresenius (F80A and 200A) and Gambro (17R) manufacture using the dead-end filtration method. Two hundred and seventeen, predominantly used but some new, dialyzers were evaluated. Water permeability of the used, but not the new, dialyzers fell abruptly and dramatically with reprocessing. The permeability fell almost 70% in the F80A dialyzer after three reprocessing procedures with similar, but somewhat slower declines, seen in the other two dialyzers. We conclude that there is a decline in water permeability seen in Renalin reprocessed dialyzers. This factor and the associated change in solute clearance and ultrafiltration characteristics should be considered in assessing the effectiveness of dialyzer reprocessing.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17377505     DOI: 10.1038/sj.ki.5002212

Source DB:  PubMed          Journal:  Kidney Int        ISSN: 0085-2538            Impact factor:   10.612


  3 in total

1.  Dialyzer reuse with peracetic acid does not impact patient mortality.

Authors:  T Christopher Bond; Allen R Nissenson; Mahesh Krishnan; Steven M Wilson; Tracy Mayne
Journal:  Clin J Am Soc Nephrol       Date:  2011-05-12       Impact factor: 8.237

2.  Asymmetric triacetate membrane keeps high water flux during ultrafiltration: in vitro study.

Authors:  Yoshihiro Tange; Shingo Takesawa; Shigenori Yoshitake
Journal:  J Artif Organs       Date:  2017-07-06       Impact factor: 1.731

3.  Abandoning peracetic acid-based dialyzer reuse is associated with improved survival.

Authors:  Eduardo Lacson; Weiling Wang; Ann Mooney; Norma Ofsthun; J Michael Lazarus; Raymond M Hakim
Journal:  Clin J Am Soc Nephrol       Date:  2010-10-14       Impact factor: 8.237