Literature DB >> 17377102

The effect of albumin concentration on plasma sodium and chloride measurements in critically ill patients.

David A Story1, Hiroshi Morimatsu, Moritoki Egi, Rinaldo Bellomo.   

Abstract

BACKGROUND: We tested the hypothesis that the difference between indirect and direct sodium assays would be related to the plasma albumin concentration. Further, we proposed that differences between indirect and direct chloride assays might be explained by interference from other plasma constituents, particularly bicarbonate, and possibly albumin.
METHODS: We studied 300 critically ill patients at the time of admission to the intensive care unit (ICU) and compared each patient's plasma sodium and chloride measurements from a central laboratory assay (indirect electrode) and an ICU blood gas machine assay (direct electrode).
RESULTS: The central laboratory sodium measurement was, on average, 2.1 mmol/L more than the ICU assay, limits of agreement 1.8-2.4 mmol/L greater, P < 0.001. The central laboratory chloride measurement was, on average, 1 mmol/L less than the ICU assay (limits of agreement 1.3-0.7 mmol/L less, P < 0.001). All correlations between the assay differences and plasma constituents were weak except for a moderately strong correlation between differences in sodium measurements and albumin. The difference in plasma sodium concentration between the assays (central laboratory - ICU) increased as the plasma concentration albumin decreased (difference = 6.2-0.16 albumin (g/L); P < 0.001, r = -0.46, r(2) = 0.22).
CONCLUSIONS: The central laboratory and ICUs assays are analytically, statistically, and clinically different for both sodium and chloride. Unless taken into account, the differences could be large enough in hypoalbuminemic populations (such as critically ill patients) to affect clinical diagnosis and decision making.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17377102     DOI: 10.1213/01.ane.0000258015.87381.61

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  13 in total

1.  Plasma sodium measurements by direct ion selective methods in laboratory and point of care may not be clinically interchangeable.

Authors:  Brigit A Weld; Thomas J Morgan; Jeffrey J Presneill; Steven Weier; David Cowley
Journal:  J Clin Monit Comput       Date:  2016-09-28       Impact factor: 2.502

Review 2.  Hyponatraemia and hypernatraemia: Disorders of Water Balance in Neurosurgery.

Authors:  Mendel Castle-Kirszbaum; Mervyn Kyi; Christopher Wright; Tony Goldschlager; R Andrew Danks; W Geoffrey Parkin
Journal:  Neurosurg Rev       Date:  2021-01-03       Impact factor: 3.042

3.  High Blood Cu/Zn Ratio is Associated with Nutritional Risk in Patients Undergoing Maintenance Hemodialysis.

Authors:  Yun Liu; Ruiying Tang; Qingdong Xu; Shilin Xu; SuJun Zuo; Jingxian Qiu; Xiaoshi Zhong; Rongshao Tan; Yan Liu
Journal:  Biol Trace Elem Res       Date:  2022-02-12       Impact factor: 4.081

4.  Potential inaccuracies in chloride measurements in patients with severe metabolic acidosis.

Authors:  Tetsuya Makiishi; Naomasa Nishimura; Keiko Yoshioka; Shinya Yamamoto; Ryuichi Mitsuhashi; Sayako Maeda; Takashi Konishi; Kunihiko Hirose
Journal:  Int J Nephrol       Date:  2012-06-12

Review 5.  Bench-to-bedside review: Chloride in critical illness.

Authors:  Nor'azim Mohd Yunos; Rinaldo Bellomo; David Story; John Kellum
Journal:  Crit Care       Date:  2010-07-08       Impact factor: 9.097

6.  Long-term changes in dysnatremia incidence in the ICU: a shift from hyponatremia to hypernatremia.

Authors:  Annemieke Oude Lansink-Hartgring; Lara Hessels; Joachim Weigel; Anne Marie G A de Smet; Diederik Gommers; Prashant V Nannan Panday; Ewout J Hoorn; Maarten W Nijsten
Journal:  Ann Intensive Care       Date:  2016-03-17       Impact factor: 6.925

7.  Are sodium and potassium results on arterial blood gas analyzer equivalent to those on electrolyte analyzer?

Authors:  Shalini Gupta; Ashwani K Gupta; Kamaljit Singh; Minni Verma
Journal:  Indian J Crit Care Med       Date:  2016-04

8.  Comparison of sodium ion levels between an arterial blood gas analyzer and an autoanalyzer in preterm infants admitted to the neonatal intensive care unit: a retrospective study.

Authors:  Hyunho Kim; Jin Kyu Kim; Soo Chul Cho
Journal:  BMC Pediatr       Date:  2016-07-20       Impact factor: 2.125

9.  Impact of chloride and strong ion difference on ICU and hospital mortality in a mixed intensive care population.

Authors:  Niels Van Regenmortel; Walter Verbrugghe; Tim Van den Wyngaert; Philippe G Jorens
Journal:  Ann Intensive Care       Date:  2016-09-17       Impact factor: 6.925

Review 10.  Molecular and physical technologies for monitoring fluid and electrolyte imbalance: A focus on cancer population.

Authors:  Devasier Bennet; Yasaman Khorsandian; Jody Pelusi; Amy Mirabella; Patrick Pirrotte; Frederic Zenhausern
Journal:  Clin Transl Med       Date:  2021-06
View more