| Literature DB >> 17375974 |
Kenneth E Ekstrand1, William H Hinson, William Kearns, J Daniel Bourland, Allan F deGuzman, Kevin P McMullen, Volker W Stieber.
Abstract
Because of geometrical limitations in the helmet of the Leksell Gamma Knife(Elekta Corp., Atlanta, GA, USA) certain regions within the cranium cannot be targeted for treatment. We describe a method by which lesions in these regions can be treated with the Varian-Zmed stereotactic radiosurgery system utilizing an infrared optical positioning system attached to a Leksell head frame. We have measured the accuracy of the optical tracking system using a phantom attached to a Leksell frame and have determined that the system can target a linear accelerator radiosurgery beam to an accuracy of within 1 millimeter.Mesh:
Year: 2007 PMID: 17375974 DOI: 10.1177/153303460700600208
Source DB: PubMed Journal: Technol Cancer Res Treat ISSN: 1533-0338