Literature DB >> 1736696

The accuracy of train-of-four monitoring at varying stimulating currents.

H S Helbo-Hansen1, U Bang, H K Nielsen, L T Skovgaard.   

Abstract

Train-of-four (TOF) monitoring of neuromuscular block at submaximal current stimulation has been recommended because it is less painful than monitoring at supramaximal stimulation. The measurement error, however, when using submaximal stimulation has not as yet been fully elucidated. The authors therefore quantified the measurement error of TOF monitoring at low currents near the initial threshold for stimulation (ITS) by assessing precision (the difference between duplicated TOF ratios measured at the same current) and accuracy (the difference between TOF ratios at lesser currents and the TOF ratio at 58 mA, which served as the reference standard method). A stable neuromuscular block at a TOF ratio of 0.40 and subsequently 0.70 was established in 20 patients using a continuous infusion of atracurium. The ulnar nerve was stimulated at the wrist using TOF stimulation, and the evoked twitch response of the adductor pollicis muscle was measured with a force-displacement transducer. The stimulating current was varied in 5-mA steps between ITS and 55 mA. Whereas there were no statistical significant differences between the mean TOF ratios at the different currents, the standard deviation increased with decreasing currents (P less than 0.01 at currents less than 40 mA). The precision was acceptable except at ITS, where the limits of precision (mean intracurrent difference in TOF ratio +/- two standard deviations) exceeded 0.05, which was the chosen acceptable difference. The standard deviation of the accuracy was significantly increased at currents less than or equal to ITS + 25 mA (P less than 0.01).(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1736696     DOI: 10.1097/00000542-199202000-00007

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  9 in total

1.  Comparison of tactile and mechanomyographical assessment of response to double burst and train-of-four stimulation during moderate and profound neuromuscular blockade.

Authors:  H Kirkegaard-Nielsen; H S Helbo-Hansen; I K Severinsen; P Lindholm; H S Pedersen; M B Schmidt
Journal:  Can J Anaesth       Date:  1995-01       Impact factor: 5.063

2.  Assessment of double-burst monitoring at 10 mA above threshold current.

Authors:  D G Silverman; S J Brull
Journal:  Can J Anaesth       Date:  1993-06       Impact factor: 5.063

Review 3.  [Neuromuscular monitoring in patients with neuromuscular diseases. Options and needs].

Authors:  A Beloiartsev; S Gableske; M Hübler
Journal:  Anaesthesist       Date:  2009-07       Impact factor: 1.041

4.  Monitoring masseter muscle evoked responses enables faster tracheal intubation.

Authors:  Osamu Kitajima; Takahiro Suzuki; Naoto Watanabe; Takeshi Maeda; Yoshikazu Noda; Shigeru Saeki; Setsuro Ogawa
Journal:  J Anesth       Date:  2010-01-21       Impact factor: 2.078

5.  New equipment for neuromuscular transmission monitoring: a comparison of the TOF-Guard with the Myograph 2000.

Authors:  H Kirkegaard-Nielsen; H S Helbo-Hansen; P Lindholm; H S Pedersen; I K Severinsen; M B Schmidt
Journal:  J Clin Monit Comput       Date:  1998-01       Impact factor: 2.502

6.  A comparison of dynamic and isometric force sensors for train-of-four measurement using submaximal stimulation current.

Authors:  S E Kern; J O Johnson; D R Westenskow; J A Orr
Journal:  J Clin Monit       Date:  1995-01

7.  Clinical assessment and train-of-four measurements in critically ill patients treated with recommended doses of cisatracurium or atracurium for neuromuscular blockade: a prospective descriptive study.

Authors:  Pierre Bouju; Jean-Marc Tadié; Nicolas Barbarot; Julien Letheulle; Fabrice Uhel; Pierre Fillatre; Guillaume Grillet; Angélique Goepp; Yves Le Tulzo; Arnaud Gacouin
Journal:  Ann Intensive Care       Date:  2017-01-19       Impact factor: 6.925

8.  Subparalyzing Doses of Rocuronium Reduce Muscular Endurance without Detectable Effect on Single Twitch Height in Awake Subjects.

Authors:  Jan Gelberg; Peter Bentzer; David Grubb
Journal:  Anesthesiol Res Pract       Date:  2019-05-02

9.  Optimised reversal without train-of-four monitoring versus reversal using quantitative train-of-four monitoring: An equivalence study.

Authors:  Ardyan Wardhana; Juni Kurniawaty; Yusmein Uyun
Journal:  Indian J Anaesth       Date:  2019-05
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.