Literature DB >> 17353654

The Health and Life device (HL888HA): accuracy assessment in an adult population according to the British Hypertension Society protocol.

Annemarie de Greeff1, Andrew H Shennan.   

Abstract

OBJECTIVE: Over the last two decades, concerns over mercury toxicity have given rise to the continuous release of new and improved automated blood pressure measurement devices as a replacement for the mercury sphygmomanometer. Out-of-office blood pressure monitoring has been shown to have various advantages, including a reduced white-coat effect and observer error, as well as better prediction of cardiovascular risk/complications. We assessed the accuracy of the Health and Life device (HL888HA) (Health and Life Co. Ltd., Taipei, Taiwan) in an adult population according to the British Hypertension Society protocol.
METHODS: Local ethics committee approval was obtained and participants were recruited from a large teaching hospital. Nine blood pressure measurements were taken from each participant, alternating between a mercury sphygmomanometer and the test device. Data from 85 participants who fulfilled the British Hypertension Society criteria were analysed according to the protocol guidelines.
RESULTS: The device achieved the highest possible grade (A/A) according to the British Hypertension Society protocol. The mean difference and SD of 0.5 (7.3) mmHg for systolic and -2.1 (7.8) mmHg for diastolic pressure, which complies with the standard of the American Association for the Advancement of Medical Instrumentation.
CONCLUSION: The Health and Life device (HL888HA) can be recommended for use in an adult population according to the criteria of the British Hypertension Society protocol.

Entities:  

Mesh:

Year:  2007        PMID: 17353654     DOI: 10.1097/MBP.0b013e32809efa64

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  1 in total

1.  Validation of the Oregon Scientific BPU 330 for self-monitoring of blood pressure according to the International Protocol.

Authors:  Li Li; XinYu Zhang; ChunHong Yan; QingXiang Liang
Journal:  Vasc Health Risk Manag       Date:  2008
  1 in total