Electrical stimulation of the epidural space using a catheter with a removable stylet.

Roshanak Charghi1, Sher Yi Chan, Kenneth J Kardash, Roderick J Finlayson, De Q H Tran.   

Abstract

BACKGROUND AND OBJECTIVES: Electrical stimulation can be used to verify the location of epidural catheters. With the traditional technique, the latter must be primed with saline to allow for electrical conduction: any air lock will, thus, hinder the flow of current. Therefore, we set out to explore an alternative mode of stimulation by use of a catheter containing a removable stylet. This study examines the reliability of this new technique.
METHODS: In 71 patients undergoing surgery, a thoracic epidural catheter was inserted preoperatively. Loss of resistance was used to identify the epidural space. The TheraCath was introduced 5 cm into the space and connected to a neurostimulator via a 2-headed alligator clip. The intensity, pulse width, and level of myotomal contractions were recorded upon stimulation of the epidural space. A bolus of lidocaine was then injected and the patient assessed for sensory block to ice.
RESULTS: The failure rate in proper epidural catheter placement was 8.5%. Epidural stimulation yielded a mean threshold of 1.90 +/- 1.80 nanocoulombs. A total of 37 catheters produced a unilateral muscular response; however, block to ice was bilateral. When compared with the ice test, the sensitivity, specificity, positive predictive value, and negative predictive value of epidural stimulation with the TheraCath were 92%, 83%, 98%, and 50%, respectively.
CONCLUSION: We conclude that the TheraCath, because of the removable stylet, provides effective electrostimulation of the epidural space. Its use was simple and devoid of complications. Nonetheless, further studies are required before implementing its routine use in clinical settings.

Entities:  

Mesh:

Year:  2007        PMID: 17350527     DOI: 10.1016/j.rapm.2006.10.006

Source DB:  PubMed          Journal:  Reg Anesth Pain Med        ISSN: 1098-7339            Impact factor:   6.288


  4 in total

1.  Strategic placement of epidural catheter.

Authors:  Jae-Hang Shim
Journal:  Korean J Anesthesiol       Date:  2012-09-14

2.  Efficacy of electrical stimulation on epidural anesthesia for cesarean section: a randomized controlled trial.

Authors:  Young Sung Kim; Hyo Sung Kim; Hyerim Jeong; Chung Hun Lee; Mi Kyoung Lee; Sang Sik Choi
Journal:  BMC Anesthesiol       Date:  2020-06-10       Impact factor: 2.217

3.  A randomized comparison of loss of resistance versus loss of resistance plus electrical stimulation: effect on success of thoracic epidural placement.

Authors:  Sean Wayne Dobson; Robert Stephen Weller; Christopher Edwards; James David Turner; Jonathan Douglas Jaffe; Jon Wellington Reynolds; Daryl Steven Henshaw
Journal:  BMC Anesthesiol       Date:  2022-02-09       Impact factor: 2.217

4.  Minimum current requirement for confirming the localization of an epiradicular catheter placement.

Authors:  Ji Seon Jeong; Jae Chol Shim; Jae Hang Shim; Dong Won Kim; Min Serk Kang
Journal:  Korean J Anesthesiol       Date:  2012-09-14
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.