On the dose to a moving target while employing different IMRT delivery mechanisms.

Eric D Ehler1, Benjamin E Nelms, Wolfgang A Tomé.   

Abstract

BACKGROUND AND PURPOSES: To compare the temporal uniformity in dose delivered to a moving target for various intensity modulation radiotherapy (IMRT) modalities: solid intensity modulator (SIM), segmented multi-leaf collimator (SMLC), and dynamic multi-leaf collimator (DMLC).
MATERIALS AND METHODS: Two separate four-dimensional computed tomography data sets were obtained. Tumor motion kernels and motion envelopes were determined from composite positions of the tumor in various phases of the breathing cycle. Treatment plans were created for an unmodulated open field, SIM, SMLC, and DMLC. The motion envelope was treated as a static target volume. A robotic apparatus equipped with a diode array simulated the tumor motion in the plane of the beam's eye view (BEV). Radiation was delivered to the moving target over ten trials for each modality. The average coefficient of variation (CV) was determined for each beam angle.
RESULTS: The CV ranged from 0.09% to 0.15%, 0.23% to 3.14%, 1.14% to 5.51%, and 3.83% to 8.25% for the unmodulated open field, SIM, SMLC, and DMLC modalities, respectively. With gating, the CV was 0.23% to 2.31%, 0.31% to 2.97%, and 0.7% to 4.67% for SIM, SMLC, and DMLC, respectively.
CONCLUSION: SIM consistently provided the most temporally uniform dose to the moving target while DMLC provided the least. The SMLC and DMLC CV improved with gated delivery.

Entities:  

Mesh:

Year:  2007        PMID: 17350124     DOI: 10.1016/j.radonc.2007.02.007

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  7 in total

1.  Dosimetric evaluation of compensator intensity modulation-based stereotactic body radiotherapy for Stage I non-small-cell lung cancer.

Authors:  Y Tajima; H Nakayama; T Itonaga; S Shiraishi; M Okubo; R Mikami; S Sugahara; K Tokuuye
Journal:  Br J Radiol       Date:  2015-05-21       Impact factor: 3.039

2.  Evaluation of a fast method of EPID-based dosimetry for intensity-modulated radiation therapy.

Authors:  Benjamin E Nelms; Karl H Rasmussen; Wolfgang A Tome
Journal:  J Appl Clin Med Phys       Date:  2010-04-19       Impact factor: 2.102

3.  Evaluation of tumor motion effects on dose distribution for hypofractionated intensity-modulated radiotherapy of non-small-cell lung cancer.

Authors:  Hyejoo Kang; Ellen D Yorke; Jie Yang; Chen-Shou Chui; Kenneth E Rosenzweig; Howard I Amols
Journal:  J Appl Clin Med Phys       Date:  2010-06-08       Impact factor: 2.102

4.  3D dosimetric validation of ultrasound-guided radiotherapy with a dynamically deformable abdominal phantom.

Authors:  Charles K Matrosic; Wesley Culberson; Andrew Shepard; Sydney Jupitz; Bryan Bednarz
Journal:  Phys Med       Date:  2021-04-23       Impact factor: 2.685

5.  A comparison of the dosimetric effects of intrafraction motion on step-and-shoot, compensator, and helical tomotherapy-based IMRT.

Authors:  Ben J Waghorn; Robert J Staton; Justin M Rineer; Sanford L Meeks; Katja Langen
Journal:  J Appl Clin Med Phys       Date:  2013-05-06       Impact factor: 2.102

6.  Quantitative analysis of brass compensators for commissioning of the Pinnacle planning system for IMRT.

Authors:  Larry L Gates; David J Gladstone
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

7.  A margin-based analysis of the dosimetric impact of motion on step-and-shoot IMRT lung plans.

Authors:  Benjamin J Waghorn; Amish P Shah; Justin M Rineer; Katja M Langen; Sanford L Meeks
Journal:  Radiat Oncol       Date:  2014-02-05       Impact factor: 3.481

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.