MR compatibility of EEG scalp electrodes at 4 tesla.

Todd K Stevens1, John R Ives, L Martyn Klassen, Robert Bartha.   

Abstract

PURPOSE: To design and apply a method to quantitatively evaluate the MR compatibility of electroencephalographic (EEG) scalp electrodes based on pulse sequence-independent metrics.
MATERIALS AND METHODS: Three types of electrodes (constructed primarily of brass, silver, and conductive plastic, respectively) were tested. B0 field distortions, B1 shielding, and heat induction was measured in adjacent agarose and oil phantoms at 4 T. B0 field maps were corrected for distortions caused by the measurement apparatus and passive shim heating, and projections perpendicular to the surfaces of the electrodes were fit, generating cubic coefficients representing the electrode distortion severity. Signal loss in T2-weighted images was used to determine B1 shielding by the electrodes. Temperature measurements were recorded during the application of a high-power pulse sequence.
RESULTS: Significantly different B0 distortions were observed in the three types of electrodes. The B1 shielding detected in all three electrodes is minimal for most human MRI, and no significant heating was detected in the electrodes or adjacent phantom.
CONCLUSION: The three types of electrodes were successfully differentiated in terms of MR compatibility based on pulse sequence-independent B0 field distortions. Copyright (c) 2007 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2007        PMID: 17345636     DOI: 10.1002/jmri.20872

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  6 in total

1.  Influence of dense-array EEG cap on fMRI signal.

Authors:  Qingfei Luo; Gary H Glover
Journal:  Magn Reson Med       Date:  2011-12-09       Impact factor: 4.668

2.  The utility of conductive plastic electrodes in prolonged ICU EEG monitoring.

Authors:  Rohit R Das; Brendan P Lucey; Sherry H-Y Chou; Patricio S Espinosa; Amir A Zamani; Barbara A Dworetzky; Edward B Bromfield; Jong Woo Lee
Journal:  Neurocrit Care       Date:  2008-09-20       Impact factor: 3.210

3.  An EEG (electroencephalogram) recording system with carbon wire electrodes for simultaneous EEG-fMRI (functional magnetic resonance imaging) recording.

Authors:  Michiro Negishi; Mark Abildgaard; Ilan Laufer; Terry Nixon; Robert Todd Constable
Journal:  J Neurosci Methods       Date:  2008-06-07       Impact factor: 2.390

Review 4.  EEG-Informed fMRI: A Review of Data Analysis Methods.

Authors:  Rodolfo Abreu; Alberto Leal; Patrícia Figueiredo
Journal:  Front Hum Neurosci       Date:  2018-02-06       Impact factor: 3.169

5.  Segmenting electroencephalography wires reduces radiofrequency shielding artifacts in simultaneous electroencephalography and functional magnetic resonance imaging at 7 T.

Authors:  Thanh Phong Lê; Rolf Gruetter; João Jorge; Özlem Ipek
Journal:  Magn Reson Med       Date:  2022-05-16       Impact factor: 3.737

Review 6.  Simultaneous EEG-fMRI: What Have We Learned and What Does the Future Hold?

Authors:  Tracy Warbrick
Journal:  Sensors (Basel)       Date:  2022-03-15       Impact factor: 3.576

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.