Literature DB >> 17341025

Determination of formoterol in rat plasma by liquid chromatography-electrospray ionisation mass spectrometry.

I Kakubari1, H Dejima, K Miura, Y Koga, H Mizu, T Takayasu, H Yamauchi, S Takayama, K Takayama.   

Abstract

A sensitive method for the determination of formoterol in rat plasma is described, using high performance liquid chromatographic separation with tandem mass spectrometry. Samples were purified using liquid-liquid extraction and separated on CAPCELL PAK C18 UG120 (2.0 x 150 mm) with a mobile phase consisting of a mixture of methanol- 50 mM ammonium hydrogen carbonate (1:1 v/v). Detection was performed with a TSQ 7000 mass spectrometer using positive ion electrospray ionisation, monitoring the shift from precursor ions for formoterol at m/z 344.9 to product ions of m/z 121.0. The limit of quantitation of the method was found to be 0.1 ng/ml, when using 0.1 ml plasma. Plasma concentrations of formoterol could be quantified from 0.15 to 7.01 ng/ml, allowing the analysis of samples up to 32 h after a single oral dose of formoterol fumarate (0.25 mg) to rats.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17341025

Source DB:  PubMed          Journal:  Pharmazie        ISSN: 0031-7144            Impact factor:   1.267