F Trompier1, S Delacroix, I Vabre, F Joussard, J Proust.
Abstract
Following modifications on the beam line at the Orsay Protontherapy Center, dose measurements were performed in order to make a dose map in the treatment rooms and in the delimited radiation-controlled area around beam line. Measurements were performed using tissue-equivalent proportional counters and rem-counters. Analysis of TEPC single event measurements showed that high LET components (>10 keV.microm(-1)) represent 90 to 99% of total dose equivalent in the treatment rooms and 50 to 90% in the controlled area and quality factors range, respectively between 2 and 15. A fast neutron component was identified in the treatment rooms, where dose equivalent rate varied between few microSv.h(-1) to some dozen of mSv.h(-1). In high-energy radiation field rem-counters underestimated TEPC values for neutron component. The variation between instruments response according to the location is linked to energetic spectrum variations and instrument characteristics.Mesh:
Substances:
Year: 2007 PMID: 17337734 DOI: 10.1093/rpd/ncm154
Source DB: PubMed Journal: Radiat Prot Dosimetry ISSN: 0144-8420 Impact factor: 0.972