Literature DB >> 17323355

Beating-heart percutaneous mitral valve repair using a transcatheter endovascular suturing device in an animal model.

Tasneem Z Naqvi1, Maurice Buchbinder, David Zarbatany, Julie Logan, Mia Molloy, Glenn Balke, Robert Ainsworth, John G Webb, Ottavio Alfieri, Francesco Maisano.   

Abstract

BACKGROUND: The edge-to-edge (Alfieri) technique for mitral valve repair is a versatile method of treating mitral insufficiency. Because of its simplicity, it has been applied in minimally invasive surgery, and recently, in the design of endovascular closed-heart devices. AIM: The purpose of this study was to evaluate the acute in-vivo safety and feasibility of a novel percutaneous mitral valve repair system based on Alfieri technique in an animal model.
METHODS: Under general anesthesia, 11 pigs (90-100 kgs), underwent percutaneous Alfeiri procedure. The right femoral vein was punctured and the mitral valve was approached via a standard transeptal puncture. Combined intracardiac echo and fluoroscopic guidance was used. The procedure included: the positioning of a guide catheter for multiple access to the left atrium and for directing devices; the use of a therapy device to capture the free edge of the mitral valve leaflets using vacuum, and to deliver the suture to the valve and finally the fixation with a Nitinol suture clip, and trimming of the suture with a fastener catheter.
RESULTS: Leaflet capture, suture placement, and suture-clip deployment was successful in all 11 animals. There were no acute cardiac or access site complications. Procedural time (from wire in left atrium to completion of the procedure was 18 +/- 9 min (range 9-30 min). Blood loss was 67 +/- 44 ml (range 0-125 ml). A double orifice configuration was visible by echocardiography at the end of the procedure in all animals.
CONCLUSION: This acute animal study demonstrated the feasibility of a beating heart percutaneous Alfieri procedure in a non-diseased porcine valve using an endovascular suturing device to safely access the mitral valve, place a stitch through the mitral valve leaflets, and deploy a suture-clip that reproduces the surgical technique. Clinical application of this device in humans needs to be evaluated. (c) 2007 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2007        PMID: 17323355     DOI: 10.1002/ccd.21029

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  7 in total

Review 1.  Percutaneous approaches to mitral valve regurgitation.

Authors:  S C Bertog; J Franke; D H Steinberg; N Wunderlich; H Sievert
Journal:  Herz       Date:  2012-03       Impact factor: 1.443

2.  Quantification of biomechanical interaction of transcatheter aortic valve stent deployed in porcine and ovine hearts.

Authors:  Joseph Mummert; Eric Sirois; Wei Sun
Journal:  Ann Biomed Eng       Date:  2012-11-17       Impact factor: 3.934

3.  Mitral cerclage annuloplasty, a novel transcatheter treatment for secondary mitral valve regurgitation: initial results in swine.

Authors:  June-Hong Kim; Ozgur Kocaturk; Cengizhan Ozturk; Anthony Z Faranesh; Merdim Sonmez; Smita Sampath; Christina E Saikus; Ann H Kim; Venkatesh K Raman; J Andrew Derbyshire; William H Schenke; Victor J Wright; Colin Berry; Elliot R McVeigh; Robert J Lederman
Journal:  J Am Coll Cardiol       Date:  2009-08-11       Impact factor: 24.094

Review 4.  Clinical trial experience with the MitraClip catheter based mitral valve repair system.

Authors:  Francesco Maisano; Cosmo Godino; Andrea Giacomini; Paolo Denti; Iryna Arendar; Nicola Buzzatti; Giovanni La Canna; Ottavio Alfieri; Antonio Colombo
Journal:  Int J Cardiovasc Imaging       Date:  2011-04-19       Impact factor: 2.357

Review 5.  Percutaneous and off-pump treatments for functional mitral regurgitation.

Authors:  Kiyotaka Fukamachi
Journal:  J Artif Organs       Date:  2008-04-15       Impact factor: 1.731

6.  Quantification of structural compliance of aged human and porcine aortic root tissues.

Authors:  Kewei Li; Qian Wang; Thuy Pham; Wei Sun
Journal:  J Biomed Mater Res A       Date:  2013-09-02       Impact factor: 4.396

Review 7.  Complications Following MitraClip Implantation.

Authors:  Katharina Schnitzler; Michaela Hell; Martin Geyer; Felix Kreidel; Thomas Münzel; Ralph Stephan von Bardeleben
Journal:  Curr Cardiol Rep       Date:  2021-08-13       Impact factor: 2.931

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.