| Literature DB >> 1732265 |
N V Wilson1, S K Das, V V Kakkar, H D Maurice, J G Smibert, E M Thomas, J E Nixon.
Abstract
We performed a prospective randomised controlled trial of a new mechanical method of prophylaxis for venous thrombo-embolism in 60 patients undergoing knee replacement surgery. The method uses the A-V Impulse System to produce cyclical compression of the venous reservoir of the foot. The overall incidence of deep-vein thrombosis was 68.7% in patients receiving no prophylaxis and 50% in those using the device. The difference was not significant. There was, however, a reduction of the extent of thrombosis in the treated group. There were 13 major calf-vein thrombi and six proximal-vein thrombi in the control group compared with only five major calf-vein thrombi in the treated group. This difference was significant (p = 0.014). No patient developed clinical features of a pulmonary embolism.Entities:
Mesh:
Year: 1992 PMID: 1732265
Source DB: PubMed Journal: J Bone Joint Surg Br ISSN: 0301-620X