Literature DB >> 17309877

Uncertainties in determination of 131I activity in thyroid gland.

Jakub Ośko1, Natalia Golnik, Tomasz Pliszczynski.   

Abstract

The values of the iodine 131I uptake determined from the measurements with a NaI scintillation counter might be significantly influenced by the thyroid position in the human neck. It is shown that the ratio of the counting rate in the energy peak of 364 keV to the counting rate in the Compton scattering band can be used for the determination of the effective depth of the thyroid. The uncertainties of the standard method are discussed basing on results of calibration measurements. The investigations of 95 patients with different thyroid diseases showed that the measured iodine spectrum was considerably different from the standard for about 20% of them, indicating possible high uncertainties of the standard measurements, which assume a fixed thyroid depth of 20 mm.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17309877     DOI: 10.1093/rpd/ncm158

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

1.  Variation of counting efficiency in determination of 131I activity in the thyroid gland as a result of its position relative to the detector.

Authors:  R Kierepko; P Janowski; M Grochowska
Journal:  J Radioanal Nucl Chem       Date:  2014-03-09       Impact factor: 1.371

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.