Literature DB >> 17309740

The influence of irrigating solutions on the accuracy of the electronic apex locator facility in the Tri Auto ZX handpiece.

A Erdemir1, A U Eldeniz, H Ari, S Belli, T Esener.   

Abstract

AIM: To determine the influence of various irrigating solutions on the accuracy of the electronic apex locator facility in the Tri Auto ZX handpiece.
METHODOLOGY: One hundred and forty teeth with single canals and mature apices, scheduled for extraction for either periodontal or prosthetic reasons in 76 patients were used. Following informed written consent local anaesthesia was administered, access cavities were prepared and pulp tissue removed. The teeth were then randomly divided into seven groups according to the irrigating solutions used. The root canal length measurements were completed using the Tri Auto ZX handpiece with automatic reverse function in the presence of one or other of the following solutions: 0.9% saline, 2.5% NaOCl, 3% H(2)O(2), 0.2% chlorhexidine, 17% EDTA, Ultracaine D-S or in the absence of an irrigating solution (control). Files were immobilized in the access cavity with composite resin. After extraction, the apical regions of the teeth were exposed and the file tips examined under a stereomicroscope. Distances between the file tips and the apical constriction were measured (mm) and analysed using a one-way anova and post hoc Tukey test.
RESULTS: Mean distances from the apical constriction to the file tip were longer in the 0.9% saline group (P<0.05). There was no statistically significant difference on file tip position between the other solutions.
CONCLUSIONS: Tri Auto ZX gave reliable results with all irrigating solutions apart from in the presence of 0.9% saline.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17309740     DOI: 10.1111/j.1365-2591.2007.01234.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  7 in total

1.  Accuracy of Root ZX in teeth with simulated root perforation in the presence of gel or liquid type endodontic irrigant.

Authors:  Hyeong-Soon Shin; Won-Kyung Yang; Mi-Ri Kim; Hyun-Jung Ko; Kyung-Mo Cho; Se-Hee Park; Jin-Woo Kim
Journal:  Restor Dent Endod       Date:  2012-08-29

2.  Calcium hydroxide dressing residues after different removal techniques affect the accuracy of Root-ZX apex locator.

Authors:  Emel Uzunoglu; Ayhan Eymirli; Mehmet Özgür Uyanik; Semra Çalt; Emre Nagas
Journal:  Restor Dent Endod       Date:  2014-11-05

3.  The accuracy of the auto-stop function of different endodontic devices in detecting the apical constriction.

Authors:  David W Christofzik; Andreas Bartols; Mahmoud Khaled; Birte Größner-Schreiber; Christof E Dörfer
Journal:  BMC Oral Health       Date:  2017-11-29       Impact factor: 2.757

4.  Influence of calcium hydroxide residues after using different irrigants on the accuracy of two electronic apex locators: An in vitro study.

Authors:  Nooshin Sadat Shojaee; Zahra Zaeri; Mohammad Mehdi Shokouhi; Fereshteh Sobhnamayan; Alireza Adl
Journal:  Dent Res J (Isfahan)       Date:  2020-01-21

5.  Accuracy of two electronic apex locators in locating root perforations in curved canals in dry and wet conditions: A comparative in vitro study.

Authors:  Monisha Parshotam Khatri; Sheetal B Ghivari; Madhu Pujar; Reshma Faras; Pallavi Gopeshetti; Amulya Vanti
Journal:  Dent Res J (Isfahan)       Date:  2019-11-12

6.  Effects of dissolving solutions on the accuracy of an electronic apex locator-integrated endodontic handpiece.

Authors:  Yakup Ustun; Ozgur Uzun; Ozgur Er; Murat Maden; Fatma Yalpı; Burhan Can Canakci
Journal:  ScientificWorldJournal       Date:  2013-11-26

7.  The influence of two forms of chlorhexidine on the accuracy of contemporary electronic apex locators.

Authors:  Ewa Marek; Ryta Łagocka; Katarzyna Kot; Krzysztof Woźniak; Mariusz Lipski
Journal:  BMC Oral Health       Date:  2019-12-31       Impact factor: 2.757

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.