Literature DB >> 17302723

Thresholds and complications with right ventricular septal pacing compared to apical pacing.

Haran Burri1, Henri Sunthorn, Pierre-André Dorsaz, Isabelle Viera, Dipen Shah.   

Abstract

BACKGROUND: Right ventricular septal pacing has been proposed as an alternative to apical pacing. However, data concerning thresholds and requirement for lead repositioning with this technique are scant.
METHODS: We reviewed data at implantation and follow-up of 362 consecutive recipients of the same model of active fixation lead (Medtronic 5076-58, Minneapolis, MN, USA) to avoid differences due to lead characteristics. Patients were divided into two groups according to whether the lead was positioned on the interventricular septum (n = 157) or at the right ventricular apex (n = 205). Thresholds, lead impedance, and requirement for lead repositioning were compared between groups at implantation and follow-up.
RESULTS: There were no differences between the septal and apical groups in sensing and pacing thresholds or lead impedance, either at implantation or during a 24-month follow-up. In the septal group, the lead had to be repositioned in four patients (2.5%) due to lead dislodgement in two patients, acute threshold rise in one patient, and pericardial effusion in another patient (the lead had unintentionally been positioned on the anterior free wall in these last two patients). In the apical group, the lead had to be repositioned in eight patients (3.9%, P = 0.56) due to lead dislodgement in three patients and acute threshold rise in five patients.
CONCLUSIONS: Acute and chronic thresholds associated with septal pacing are similar to those observed with apical pacing, and risk of lead dislodgement is low. However, multiple radioscopic views must be used to avoid inadvertent positioning of the lead on the anterior free wall.

Entities:  

Mesh:

Year:  2007        PMID: 17302723     DOI: 10.1111/j.1540-8159.2007.00610.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  7 in total

1.  Comparison of left ventricular systolic function and mechanical dyssynchrony using equilibrium radionuclide angiography in patients with right ventricular outflow tract versus right ventricular apical pacing: A prospective single-center study.

Authors:  Harmandeep Singh; Chetan D Patel; Gautam Sharma; Nitish Naik
Journal:  J Nucl Cardiol       Date:  2015-01-23       Impact factor: 5.952

2.  Electrocardiographic predictors of validated right ventricular outflow tract septal pacing for correct localization of transthoracic echocardiography.

Authors:  Huiqiang Wei; Jiaojiao Tang; Dongli Chen; Qianhuan Zhang; Yuanhong Liang; Lie Liu; Shulin Wu; Chunying Lin; Zhiming Yang; Chanjuan Chai
Journal:  Clin Cardiol       Date:  2018-03-25       Impact factor: 2.882

3.  Right ventricular pacing and sensing function in high posterior septal and apical lead placement in cardiac resynchronization therapy.

Authors:  Hm Kristiansen; T Hovstad; G Vollan; S Faerestrand
Journal:  Indian Pacing Electrophysiol J       Date:  2012-01-31

4.  Computed tomography validated right ventricular mid-septal lead implantation using right ventricular angiography.

Authors:  Jayaprakash Shenthar; Maneesh K Rai; Siva S Chakali; Vivek Pillai; Tammo Delhaas
Journal:  J Arrhythm       Date:  2021-07-11

5.  Radial left ventricular dyssynchrony by speckle tracking in apical versus non apical right ventricular pacing- evidence of dyssynchrony on medium term follow up.

Authors:  Dinesh Choudhary; Amit Kumar Chaurasia; S Mahesh Kumar; Ajeet Arulkumar; Anees Thajudeen; Narayanan Namboodiri; G Sanjay; S P Abhilash; V K Ajitkumar; Tharakan Ja
Journal:  J Cardiovasc Thorac Res       Date:  2016-03-14

6.  A randomized comparison of fluoroscopic techniques for implanting pacemaker lead on the right ventricular outflow tract septum.

Authors:  Dongli Chen; Huiqiang Wei; Jiaojiao Tang; Lie Liu; Shulin Wu; Chunying Lin; Qianhuan Zhang; Yuanhong Liang; Silin Chen
Journal:  Int J Cardiovasc Imaging       Date:  2016-01-21       Impact factor: 2.357

7.  Subclinical cardiac perforation by cardiac implantable electronic device leads detected by cardiac computed tomography.

Authors:  Yeong-Min Lim; Jae-Sun Uhm; Min Kim; In-Soo Kim; Moo-Nyun Jin; Hee Tae Yu; Tae-Hoon Kim; Hye-Jeong Lee; Young-Jin Kim; Boyoung Joung; Hui-Nam Pak; Moon-Hyoung Lee
Journal:  BMC Cardiovasc Disord       Date:  2021-07-19       Impact factor: 2.298

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.