| Literature DB >> 17302391 |
Daniel B Milligan1, Gregory J Francis, Barry J Prince, Murray J McEwan.
Abstract
The rate coefficients of the ion-molecule reactions between H3O+, NO+, O2+, and phosphine were determined using a selected ion flow tube. Using these data, the selected ion flow tube mass spectrometry (SIFT-MS) method was applied to the real-time measurement of phosphine in nitrogen without sample preparation down to concentrations in the mid parts per trillion range. This is the first reported measurement using SIFT-MS in the parts per trillion range. Linear dependencies on concentration were found from 190 ppt to the ppm range, and the limit of detection for a 10-s scan was 190 ppt (0.27 pg/mL).Entities:
Year: 2007 PMID: 17302391 DOI: 10.1021/ac0622678
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986