Noise-tolerance analysis for detection and reconstruction of absorbing inhomogeneities with diffuse optical tomography using single- and phase-correlated dual-source schemes.

B Kanmani1, R M Vasu.   

Abstract

An iterative reconstruction procedure is used to invert intensity data from both single- and phase-correlated dual-source illuminations for absorption inhomogeneities. The Jacobian for the dual source is constructed by an algebraic addition of the Jacobians estimated for the two sources separately. By numerical simulations, it is shown that the dual-source scheme performs superior to the single-source system in regard to (i) noise tolerance in data and (ii) ability to reconstruct smaller and lower contrast objects. The quality of reconstructions from single-source data, as indicated by mean-square error at convergence, is markedly poorer compared to their dual-source counterpart, when noise in data was in excess of 2%. With fixed contrast and decreasing inhomogeneity diameter, our simulations showed that, for diameters below 7 mm, the dual-source scheme has a higher percentage contrast recovery compared to the single-source scheme. Similarly, the dual-source scheme reconstructs to a higher percentage contrast recovery from lower contrast inhomogeneity, in comparison to the single-source scheme.

Mesh:

Year:  2007        PMID: 17301462     DOI: 10.1088/0031-9155/52/5/013

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  1 in total

1.  High-Speed GPU-Based Fully Three-Dimensional Diffuse Optical Tomographic System.

Authors:  Manob Jyoti Saikia; Rajan Kanhirodan; Ram Mohan Vasu
Journal:  Int J Biomed Imaging       Date:  2014-04-10
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.